VARIOUS SETTING (MENU/Log)
Indication of log and transmission
Table 9 : Display indication and the process to transmit past all date chamber temperature record
value
Description of operation
Key operated
Indication after operation
The i
ndic
at
ion o
f t
he l
og
dat
a
1
-----
Top screen is displayed
2
Press the MENU button.
MENU
The menu window is indicated. And the
“set” is highlighted.
3
Select Log by pressing Shift
keys, and press the ENTER key.
ENTER
The record value of 1-day of chamber
temperature is indicated by the graph in the
display.
4
Press the ENTER key.
ENTER
The date of the graph changes, and a graph
indicates log data on the kept past in every
1-day.
The t
ransmis
si
on of t
he
log dat
a
5
The shift key is pressed, and the
graph of a date to transmit to the PC
is indicated.
A menu window is indicated, and the log of PC
All
(
PC All
)
is highlighted.
6
The shift key is pressed, and the
graph of the all date to transmit to the
PC is indicated.
ENTER
The screen (Progress) which transmits log
data on chamber temperature of all date doing
a record to the personal computer is indicated.
7
Highlight indicates the start (Start) of
the menu window, and press the
ENTER key.
ENTER
Log data on all date (About five weeks is a
Max.) of chamber temperature are indicated
by the graph in the display.
8
A completion message (Finished.) is indicated.
The end of t
he
log da
ta
dis
pl
ay
9
If a completion message is indicated,
press the MENU button.
MENU
The menu window is indicated. And the
“Start” is highlighted.
10
A shift key is pressed, and Cancel of
the menu window is chosen, and
press the ENTER key.
ENTER
Log data on 1-day of chamber temperature are
indicated by the graph in the display.
11 Press the MENU button.
MENU
The record (PC 1D) of 1-day of the menu
window is highlighted.
12
The shift key is pressed, and made
to highlight Cancel of the menu
window, and press the ENTER key.
ENTER
Top screen is displayed
Note:
When this function is used, an optional interface board (MTR-480 only) is necessary. Contact our
sales representative or agent at the time of installation of the optional parts.
Settlement of each parameter:
With the shift key ,
,
the date is shifted. ( ; past date, ; new date)
• The settable range is between 2 and 30 minutes. The factory setting is 15 minutes. With 15
minutes log interval, the recording for about 5 weeks is available.
•
For the indication resolution of the display screen : About 1
o
C / dot and about 10 minutes/dot
• When all saved chamber temperature log value is erased, a menu button is pressed, and the log (Log) of
the menu window is chosen, and a screen is indicated when the clear (Clear) of the menu window is
chosen. All log data are erased when “OK” is chosen by the menu window.
• If log data are full, it is erased automatically in order from the old data.