9
FREEZER COMPONENTS
1.
Outer door:
To open the door, grip the door latch. On closing, lock the door latch completely.
2. Door latch:
Always lock the door latch when the outer door is closed.
3. Magnetic door gasket:
This provides a tight door seal and prevents cold air leak. Keep clean.
4. Access port (rear and bottom):
This is used for leading a cable and sensor of a measuring
equipment, or nozzle of backup cooling kit to chamber.
5. Inner door:
The operation of the inner door should be quick to minimize the temperature rise in
chamber. Lock the inner door latch completely when the outer door is closed. The inner door is
removable for cleaning or defrosting. See page 21 “Routine maintenance”.
6. Lock:
Turn clockwise to 180
o
with a key and the outer door is securely locked.
7. Control panel:
Used for temperature setting and indication of operating status is displayed on the
digital temperature indicator. See page 10 for details.
8. Caster:
4 casters are provided to facilitate moving of the cabinet. For the installation, adjust the
leveling feet so that the front 2 casters cannot contact with the floor.
9. Leveling foot:
The height of the freezer can be adjusted by this screw type foot. Keep the unit in
level at the installation.
10. Fixture (on back side):
2 fixtures serve as spacers between the cabinet and wall and also serve as
hooks to fix the unit. See page 12 “Installation”.
11. Air intake vent (grille):
Do not block this vent to keep the proper cooling performance.
12. Power switch:
This is for turning ON/OFF the power to the unit. ON – “I” OFF – “
○
”
13. Space for temperature recorder:
A temperature recorder (optional component) can be attached
here. See page 29 “Temperature recorder”.
14. Condenser filter (behind the grille):
This filter prevents the dust from accumulating on the
condenser. The dusty condenser filter may cause failure of refrigerating device. Clean the condenser
filter once a month. See page 20 “Routine maintenance” for the cleaning.
15. Battery switch:
This is a switch for a battery for power failure alarm. Normally, turn on this switch.
Be sure to turn off this switch if the freezer is not in operating for the long period. (More than 1 month)
16.
Remote alarm terminal:
This is used to notice an alarm condition of the unit to remote location.
Refer to page 18 “Remote alarm terminal”.
17. Space for optional component: