Panasonic AK-HRP1015 Firmware Update Procedure Download Page 4

 

AK-HRP1015  Firmware Update Procedure /  Mar.2021 

 

 

Page4/4 

 

Panasonic Corporation 

 

8.  When pressing the 

MENU dial C

, the update will start. 

NOTE

: Pressing the 

MENU dial B

 while 

“No (EXIT)”

 is selected, the update will be cancel. 

 

 Software update takes 

approx. 3min. 

Do not power down while updating.

 

 

9.  When the update is completed, the message 

“COMPLETE”

 is displayed. 

10. When pressing the 

MENU dial C

, ROP will reboot automatically. After reboot, confirm that the version number 

is renewed for your confirmation update was correctly done follow the item 

“ Version display method”

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3.  Confirm the version number is renewed 

 

 

Follow the procedure of “Step1”, confirm the version number is renewed.   

If it succeeded to update, the update file in the SD memory card is not need anymore. 

 

If the version update fails, turn the power to ON again after inserting SD card in which the update file is saved. 

ROP SETTING                    5 / 6 

IRIS CALIBRATION 

UPGRADE      TOP      BOTTOM 
   NO?         NO?        NO? 
 

SYSTEM VERSION       1.00-00-0.00 

 

      Confirm      
 

Do you upgrade it? 

 

 

 

 

 

No 

( ENC2 ) 

 

Yes 

( ENC3 ) 

 

Press 

MENU 

dial C 

 

ROP SETTING                    5 / 6 

IRIS CALIBRATION 

UPGRADE      TOP      BOTTOM 
   NO?          NO?        NO? 
 

SYSTEM VERSION       1.00-00-0.00 

 

      Confirm      
 

Do you upgrade it? 

 

 

 

 

 

No 

( EXIT ) 

Yes 

( ENC3 ) 

ROP SETTING                    5 / 6 

IRIS CALIBRATION 

UPGRADE      TOP      BOTTOM 
   NO?         NO?        NO? 
 

SYSTEM VERSION       1.00-00-0.00 

 

      Information      
 
ROP Upgrading… 
NOT REMOVE CARD 

 

 

ROP SETTING                    5 / 6 

IRIS CALIBRATION 

UPGRADE      TOP      BOTTOM 
   NO?         NO?        NO? 
 

SYSTEM VERSION       1.00-00-0.00 

 

      Information      
 
COMPLETE 
Push OK to reboot system. 

 

 

 

 

Yes 

( ENC3 ) 

Wait 3 minutes. 

 

Reboot 

 

Press MENU dial C 

 

Summary of Contents for AK-HRP1015

Page 1: ... button The TOP menu screen of ROP menu appears 3 Select ROP SETTING by pressing the Function button 4 Press the EXEC button to open the ROP SETTING menu 5 When pressing the Function button SYSTEM version and each Software version can be confirmed SETUP button Function button MENU 7 8 37 ROP SETTING 38 CONNECT SETTING 39 ROP IP SETTING 40 CAMERA IP SETTING 41 SD CARD STORE 42 SD CARD LOAD EXEC but...

Page 2: ...d format it NOTE SD memory card used in this unit requires to be conformed to SDTM standards Be sure to format SD memory card on this unit 3 Copy the update file 1 Accept the software usage agreement and download the update file data DV91000 zip for the AK HRP1015 firmware from the following support web site https pro av panasonic net en Support PASS Service Support website 2 Copy the file DV91000...

Page 3: ...tton to open the ROP SETTING menu 6 Select UPGRADE by pressing the Function button 7 Select YES by turning the MENU dial A and press the MENU dial A Confirm dialog appears Function button MENU dial A MENU dial B MENU dial C SETUP button EXEC button Function button MENU 7 8 37 ROP SETTING 38 CONNECT SETTING 39 ROP IP SETTING 40 CAMERA IP SETTING 41 SD CARD STORE 42 SD CARD LOAD ROP SETTING 5 6 IRIS...

Page 4: ...rm the version number is renewed If it succeeded to update the update file in the SD memory card is not need anymore If the version update fails turn the power to ON again after inserting SD card in which the update file is saved ROP SETTING 5 6 IRIS CALIBRATION UPGRADE TOP BOTTOM NO NO NO SYSTEM VERSION 1 00 00 0 00 Confirm Do you upgrade it No ENC2 Yes ENC3 Press MENU dial C ROP SETTING 5 6 IRIS...

Reviews: