For CE Mark compliance, the
Sentinel LNG
flow transmitter must meet the EMC and LVD directives.
IMPORTANT:
CE Mark compliance is required only for units intended for use in EEC countries, and is recommended
for high electrical noise areas.
B.2 EMC Compliance
In addition to the standard wiring requirements, the electrical connections must be shielded and grounded as in
Table 5 below for EMC compliance. After all the necessary electrical connections have been made, seal any unused
cable entry holes.
Note:
If the instructions in this appendix are followed, the unit will comply with the EMC Directive 2004/108/EC.
Note:
Make sure to connect the
Sentinel LNG
case to the earth ground with a grounding cable, using the external
ground screws found on either side of the enclosure
*Terminate the cable shields to the closest screw on the bus bar inside the enclosure.
Table 5: Wiring Modifications
Connection
Cable Type
Termination Modification
Transducer
Shielded cable
Terminate the shield using the cable glands (already done).
Power
Shielded cable
Terminate the shield to the case using the cable glands.
Shielding
Wires enclosed in a properly-grounded metal conduit do not require additional
shielding.
All Input/Output
Options
Shielded cable
Terminate the shield to the case.*
Summary of Contents for Sentinel LCT
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 6: ...Preface iv Sentinel LNG User s Manual...
Page 12: ...Contents x Sentinel LNG User s Manual...
Page 26: ...Chapter 1 Features and Capabilities Sentinel LNG User s Manual 14...
Page 178: ...Chapter 5 Using PanaView 166 Sentinel LNG User s Manual...
Page 184: ...Chapter 6 MODBUS Communications 172 Sentinel LNG User s Manual...
Page 202: ...Chapter 7 HART Communications 190 Sentinel LNG User s Manual...
Page 220: ...Chapter 9 Troubleshooting 208 Sentinel LNG User s Manual...
Page 222: ...Appendix A Menu Maps 210 Sentinel LNG User s Manual...
Page 230: ...Appendix A Menu Maps Sentinel LNG User s Manual 218...
Page 232: ...Appendix B CE Mark Compliance and High Noise Areas 220 Sentinel LNG User s Manual...
Page 236: ...Appendix C Data Records 224 Sentinel LNG User s Manual...
Page 240: ...Appendix C Service Record 228 Sentinel LNG User s Manual...
Page 244: ...Index 232 Sentinel LNG User s Manual...
Page 246: ...Warranty 234 Sentinel LNG User s Manual no content intended for this page...
Page 247: ......