Chapter 7. HART Communications
182
Sentinel LNG User’s Manual
7.9.1.1
Command 34 and 55 Damping Value
The variable damping value is fixed and not programmable.
7.9.1.2
Command 35 Units Restriction
Command 35 does not allow range to be set with units other than the current publishing units. If the units do not
match the selected Primary Variable units, the response "Invalid Selection" is returned. The meter does not provide
the capability to convert values between units in Command 35.
7.9.1.3
Command 41 Perform Self Test
Due to the complex nature of the SEN898SEN898 meter, the self-test mode is Restricted. However, the device status is
constantly monitored and Command 48, Read Additional Device Status, should be used to determine the current
state of the meter.
7.9.1.4
Command 43 Set Primary Variable Zero
Only the Totals variable can be set to Zero with this command. Any other variable that is not a Forward or Reverse
Total will return Invalid Selection.
7.9.1.5
Command 72 Squawk Command
The Squawk function, command 72, causes the LCD backlight to flash.
7.9.1.6
Command 52 Set Device Variable Zero
This is the same as the Device Specific command 181: Clear Totals, if the variable specified is a Totals variable. Any
totals variable specified with this command will affect all totals variables. This command does not affect variables
that are not a Totals variable. Invalid selection is returned for non-totals variables.
7.9.2 Burst Mode Commands
The
Sentinel LNG
does not support burst mode.
7.9.3 Catch Device Variable Command
The
Sentinel LNG
does not support the catch device variable mechanism.
Summary of Contents for Sentinel LCT
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 6: ...Preface iv Sentinel LNG User s Manual...
Page 12: ...Contents x Sentinel LNG User s Manual...
Page 26: ...Chapter 1 Features and Capabilities Sentinel LNG User s Manual 14...
Page 178: ...Chapter 5 Using PanaView 166 Sentinel LNG User s Manual...
Page 184: ...Chapter 6 MODBUS Communications 172 Sentinel LNG User s Manual...
Page 202: ...Chapter 7 HART Communications 190 Sentinel LNG User s Manual...
Page 220: ...Chapter 9 Troubleshooting 208 Sentinel LNG User s Manual...
Page 222: ...Appendix A Menu Maps 210 Sentinel LNG User s Manual...
Page 230: ...Appendix A Menu Maps Sentinel LNG User s Manual 218...
Page 232: ...Appendix B CE Mark Compliance and High Noise Areas 220 Sentinel LNG User s Manual...
Page 236: ...Appendix C Data Records 224 Sentinel LNG User s Manual...
Page 240: ...Appendix C Service Record 228 Sentinel LNG User s Manual...
Page 244: ...Index 232 Sentinel LNG User s Manual...
Page 246: ...Warranty 234 Sentinel LNG User s Manual no content intended for this page...
Page 247: ......