Chapter 4. Optional Display/User Interface
26
DewPro<Superscript>® MMR31 Installation & Operation Manual
4.4.5
Error Codes
During normal MMR31 operation, it is possible for the temperature channel to be in error, while the RH channel still
reads correctly. It is also possible for the RH channel to be in error, while the temperature channel still reads correctly.
If a moisture unit that depends on both the RH and temperature values (e.g., dewpoint) is selected, the error
conditions for both RH < 0% and RH > 100% will be set. MMR31 errors are indicated in the matrix (see
Table 9:
Function
Location
Description
Display the
Present
Error Code
VH 90
This location displays a number associated with an error code. If no error condition
exists, the location displays
0
. Use the instructions in this section to properly interpret the
error code.
The procedure for interpreting the
VH 90
error codes involves the conversion of numbers from decimal format into
binary format. To perform these conversions, refer to
Table 10:
Decimal Value
Binary Equivalent
0
0000
1
0001
2
0010
3
0011
4
0100
5
0101
6
0110
7
0111
8
1000
9
1001
10
1010
11
1011
12
1100
13
1101
14
1110
15
1111