Aurora H2O User’s Manual
21
Chapter 4. Programming Advanced Features
4.7.2 Setting the Decimal Format
The
Decimal Format
option determines whether a decimal [
.
] or a comma [
,
] is used as
the decimal separator. To edit the decimal format, from the
Regional Settings Menu
select
Decimal Format
and press
Enter
. The following screen appears.
Use the left and right arrow keys to select a decimal [
.
] or a comma [
,
] as the decimal
separator and press
Enter
. The screen returns to the
Display Menu
.
4.7.3Setting the Date Format
To edit the date format, from the
Regional Settings Menu
select
Date Format
and press
Enter
.
The following screen appears.
Use the left and right arrow keys to select the desired date format and press
Enter
. The
screen returns to the previous display.
4.7.4Setting the Unit System
IMPORTANT:
If the Country Code (see “Setting the Country Code” on page 20) has been set to 81 (Japan) the
Unit System menu is not available.
To select the unit system to be used for measurements, select
Unit System
and press
Enter
. The following screen appears.
Use the left and right arrow keys to select the
Unit System
desired
[SI
= only metric units
are available in the selection menus;
SI + US
= both metric and English units are
available in the selection menus] and press
Enter
. The screen returns to the
Regional
Settings Menu
.
Note:
Changing the
Unit System
setting from
SI+US
to
SI
in the above menu will
NOT
immediately change the current
Display
or
Outputs
units. This is because it is not
possible for the
Aurora H2O
to know the
SI
measurement units desired by the operator to replace any
currently displayed or output English units. However, the next time the operator selects different
Display
(see
“Selecting Primary Units” on page 8
) or
Outputs
“Selecting Output Units” on page 10
) units, only
SI
units
will be available in the menu.
4.8 User Calibration
Beginning with software revision
H2O.001.F
, the
Aurora H2O
supports a two-point
User Calibration
for use in
methane/natural gas. This is provided so that a field calibration can be easily performed to cause the
Aurora H2O
measurement to correspond to a calibration gas or other moisture device. The
User Calibration
is independent from,
and does not affect, the
Factory Calibration
. The
Aurora H2O
can be quickly switched between
User
and
Factory
calibration at any time.
Summary of Contents for Aurora H2O
Page 1: ...panametrics com 910 317 B August 2022 Aurora H2O User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Chapter x Aurora H2O User s Manual no content intended for this page...
Page 24: ...Chapter 1 Features and Capabilities 12 Aurora H2O User s Manual...
Page 44: ...Chapter 2 Installation 20 Aurora H2O User s Manual...
Page 47: ...Aurora H2O User s Manual 28...
Page 65: ...Chapter 3 Operation and General Programming 18 Aurora H2O User s Manual...
Page 96: ...Aurora H2O User s Manual 36...
Page 120: ...Chapter 5 AuroraView Interface Software 24 Aurora H2O User s Manual...
Page 140: ...Chapter 7 Troubleshooting 6 Aurora H2O User s Manual...
Page 148: ...Appendix A MODBUS RTU TCP Communications 14 Aurora H2O User s Manual...
Page 166: ...4 Aurora H2O User s Manual...
Page 168: ...Warranty 6 Aurora H2O User s Manual no content intended for this page...
Page 169: ......