Aurora H2O User’s Manual
17
Chapter 2. Installation
2.8 Making Electrical Connections (cont.)
6.
Prepare a shielded CAT5e Ethernet cable with flying leads compatible with the TIA/EIA-568 specification. For
proper operation, Aurora H2O requires connection of the TX and RX pairs, as well as the shield wire. Connect the
leads per
Figure 29
and the table below, depending on the cable used:
Figure 29: Ethernet Terminal Block
7.
For connection to a PC to interface with
AuroraView
software, you may use the supplied 704-688 cable (RS-232
w/ SUB-D-9 connector to tinned leads). Wire the cable as follows:
The as-shipped default configuration is as follows:
Lead Color
TIA/EIA 568A
TIA/EIA 568B
White/Green
TX+
RX+
Green
TX-
RX-
White/Orange
RX+
TX+
Orange
RX-
TX-
Lead Color
Function
Aurora H2O Terminal
White
TX
+
Red
RX
–
Green
Ground
RTN
BAUD Rate:
115, 200
Parity:
Even
Network ID:
1 for Port 1, 2 for Port 2
Summary of Contents for Aurora H2O
Page 1: ...panametrics com 910 317 B August 2022 Aurora H2O User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Chapter x Aurora H2O User s Manual no content intended for this page...
Page 24: ...Chapter 1 Features and Capabilities 12 Aurora H2O User s Manual...
Page 44: ...Chapter 2 Installation 20 Aurora H2O User s Manual...
Page 47: ...Aurora H2O User s Manual 28...
Page 65: ...Chapter 3 Operation and General Programming 18 Aurora H2O User s Manual...
Page 96: ...Aurora H2O User s Manual 36...
Page 120: ...Chapter 5 AuroraView Interface Software 24 Aurora H2O User s Manual...
Page 140: ...Chapter 7 Troubleshooting 6 Aurora H2O User s Manual...
Page 148: ...Appendix A MODBUS RTU TCP Communications 14 Aurora H2O User s Manual...
Page 166: ...4 Aurora H2O User s Manual...
Page 168: ...Warranty 6 Aurora H2O User s Manual no content intended for this page...
Page 169: ......