2. SPECIFICATION

Measuring Values

Humidity Sensor

Measuring range:

0...100 % Rh

Accuracy incl. hysteresis

±2.0 % Rh (0...90 % Rh)

and nonlinearity in air: 

±3.0 % Rh (90...100 % Rh)
Traceable to internal
standards, administrated by
NIST, PTB, BEV...

Temperature dependence % Rh: ±(0.022 + 0.02 x % Rh) x 

Δ

T [°C]

Δ

T = T - 20 °C

T: ±(0.0003 °C/°C)

Temperature Sensor

Working range sensing probe: 

-25...125 °C (-13...257 °F)

Outputs

Two analogue outputs for

4-20 mA Load <500 Ohm

% Saturation and Temperature: Out 1 = 0 to 100 %

Out 2 = -25 °C to 125 °C
(-13 °F to 257 °F)

General

Supply voltage:

21...28V DC

Current consumption at 24V DC:Typically 80mA

Pressure range: 

0...20 bar (0...290 psi) 
or 0...100 bar (0...1450 psi)

Housing / Protection class:

Al Si 9 Cu 3 / IP65

Electrical connection:

M12 connector

Fluid temperature range:

-40...125 °C (-40...257 °F)

Electronics temperature range: -40...80 °C (-40...176 °F)
Storage temperature range:

-40...60 °C (-40...140 °F)

Electromagnetic compatibility:

according to EN61326-1,
EN61326-2, EN61326-3.

FCC Part 15 Class B.

Industrial Environment 
ICES-003 Class B.

1. PRODUCT 

DESCRIPTION

1.1

General
The Pall WS10 Water sensor was designed for the
measurement of water content in oil. WS10 is
used for online monitoring of moisture in
lubrication or insulation oil, which is very
important for the long-term performance and
preventive maintenance of plant and machinery.
Humidity measurement in oil is similar to the
humidity measurement in the air; the actual water
content in oil can be indicated by the relative
value % RH.
% RH (actual water content expressed as a
fraction of the water content in saturated oil) and
called percent relative humidity (% RH). % RH = 0
corresponds to completely water-free oil, while %
RH = 100 indicates saturated oil. Saturation is the
point where oil holds as much water as possible
in solution, any further addition of water will result
in the presence of droplets of free water. % RH
measurement with the WS10 transmitter is based
on the outstanding long term stability and
resistance to pollution of the capacitive sensor
elements.
The WS10 transmitter has two analogue outputs,
4–20mA, for connection to customers existing
SCADA and DCS systems. The analogue outputs
represent humidity in % RH, and temperature. If
oil is cloudy due to free water contamination at
the measurement temperature, the WS10 Water
Sensor will output a signal that represents 100 %
saturation, until steps are taken to bring the water
content below the saturation point.

1.2

Dimensions in mm (inch)

M12 connector

spanner width:
27 or 24

Summary of Contents for WS10 Series

Page 1: ...WS10 WS10 Series Water Sensor service instructions IMSIWS10 EN WS10 Operating Instructions ENGLISH ...

Page 2: ... Dimensions in mm inch 4 2 SPECIFICATION 4 3 INSTALLATION 5 3 1 Installation location 5 3 2 Installing the probe directly into the process 5 4 ELECTRICAL CONNECTIONS 5 5 MAINTENANCE 5 5 1 Sensor cleaning 5 5 2 Self diagnostic and error messages 5 6 ORDERING INFORMATION 5 7 DECLARATION OF CONFORMITY 6 8 TRADEMARKS AND INTELLECTUAL PROPERTY 7 9 WARRANTY LIMITATION OF LIABILITY 7 Contents ...

Page 3: ...e time of publication WARNINGS CAUTIONS AND NOTES Care must be taken in referring to this manual to ensure adherence with all warnings cautions and important notes These carry information related to the safety of personnel and the integrity and satisfactory operation of plant ENGLISH Safety instructions General Safety Instructions Excessive mechanical loads and incorrect usage should always be avo...

Page 4: ...t in oil WS10 is used for online monitoring of moisture in lubrication or insulation oil which is very important for the long term performance and preventive maintenance of plant and machinery Humidity measurement in oil is similar to the humidity measurement in the air the actual water content in oil can be indicated by the relative value RH RH actual water content expressed as a fraction of the ...

Page 5: ...e to circulate freely around the sensing element Mount the transmitter directly into the circulating system in a pipe that is always full of oil Reservoir mounting is not ideal unless the reservoir residence time is short 3 2 Installing the probe directly into the process When installing a probe directly into the process there should be a stop valve on both sides of the point of installation This ...

Page 6: ...ry use EMC requirements Test configuration operational conditions and performance criteria for transducers with integrated or remote signal conditioning ICES 003 ClassB This Interference Causing Equipment Standard sets out the technical requirements relative to the radiated and conducted radio noise emissions from digital apparatus FCC Part 15 ClassB Covers the regulations under which an intention...

Page 7: ...r repair alteration accident or negligence in use storage transportation or handling In no event will the Seller be liable for any damages incidental consequential or otherwise whether arising out of or in connection with the manufacture packaging delivery storage use misuse or non use of any of its products or any other cause whatsoever LIABILITY Pall does not accept warranty and liability claims...

Page 8: ...r area please go to www pall com contact Because of technological developments related to the products systems and or services described herein the data and procedures are subject to change without notice Please consult your Pall representative or visit www pall com to verify that this information remains valid Pall Industrial Manufacturing 25 Harbor Park Drive Port Washington NY 11050 1 516 484 3...

Reviews: