6
For laboratory use only.
Keypad buttons on this product intended for finger touch only.
Only use recommended solutions for cleaning/sanitization.
The unit is always on standby with battery power. Avoid pressing power or pulse buttons while cleaning
pump head or loading Fluid Path.
Prolonged use in environments exceeding 30 ºC (86 ºF) may cause overheating.
The device utilizes a peristaltic pump drive which may cause a pinch point if operated with the loading
door open.
Return product in original packaging for warranty repair. Warranty valid only when used in accordance
with this User Guide.
Keypad and Symbols
MAIN POWER INDICATOR
A green LED will illuminate when the pump is plugged into the power transformer and VAC power
source. This LED will not be illuminated when operating on battery power.
POWER ON/OFF
Press the Power On/Off button to turn on the pump. Press the Power On/Off a second time to turn off
the pump.
PULSE OPERATION
The pump will operate only while the button is pressed. This feature is ideal for small fluid volumes that
require momentary filtration intervals.
FAULT INDICATOR
This symbol will illuminate yellow upon detection of a fault in the system operation. Flashing indicates
battery is running low and solid illumination indicates other faults detected. Consult the Troubleshooting
section of this User Guide.
BATTERY INDICATOR
When plugged into the power transformer and VAC power source, this symbol will indicate the battery
is charging by an illuminated blue LED. When the battery is fully charged, the blue LED will turn off.
Summary of Contents for Sentino