Pall Sentino User Manual Download Page 1

User Guide

Sentino

®

Microbiology Pump 

Product Number: 13186

For laboratory use. Not for use in a manner other than indicated.

Summary of Contents for Sentino

Page 1: ...User Guide Sentino Microbiology Pump Product Number 13186 For laboratory use Not for use in a manner other than indicated ...

Page 2: ...n 3 Product Terminology 4 Symbols Referenced 5 Safety Precautions 6 Keypad and Symbols 7 Preparing System for Use 8 Pump Operation 9 Maintenance 10 Troubleshooting 11 Service 11 Specifications 12 Disposal Instructions 12 EC Declaration of Conformity 13 2 ...

Page 3: ...need for separate manifold pump and excess tubing The Sentino Pump is designed to use peristaltic action to draw the sample through the filter and send filtrate directly to drain or waste collection Operating parameters are preset and fixed to best satisfy requirements for MF Technique described in EPA ISO and ASTM methods thereby providing conditions comparable to a standard manifold set up Order...

Page 4: ...cator Body 6 Stage with adapter guide 7 Drive tube guide 8 Product identification label 9 Power entry socket Pump Head 10 Loading door 11 Slot for drive tube Power 12 Transformer power entry plug 13 Transformer input interface 14 Power cord Fluid Path 15 Filter funnel adapter 16 Drive tube 1 3 2 4 5 6 15 14 12 11 10 16 7 9 8 13 ...

Page 5: ...all com weee to arrange for proper disposal of used equipment Safety Precautions To protect personal safety and ensure correct use of this equipment be sure you read and understand this guide before use The following terms are used DANGER To designate a condition which MAY involve mortal danger or serious injury CAUTION To designate a condition which MAY involve serious injury or destruction of th...

Page 6: ...s plugged into the power transformer and VAC power source This LED will not be illuminated when operating on battery power POWER ON OFF Press the Power On Off button to turn on the pump Press the Power On Off a second time to turn off the pump PULSE OPERATION The pump will operate only while the button is pressed This feature is ideal for small fluid volumes that require momentary filtration inter...

Page 7: ...d Path Installation The Sentino Microbiology Pump is designed to interface with Pall s disposable Fluid Path The Fluid Path will remain secure during filter funnel loading and unloading A guide on the pump housing helps position the tubing correctly inside the device without the use of any tools The device is designed for use with Pall s Fluid Path only The unit is always on standby with battery p...

Page 8: ...ss and release the power On Off button again OR Pulse or momentary operation 3a Press and hold the pulse button to perform momentary filtration 3b Release the pulse button to stop filtration 4 Perform rinses in the same manner as sample filtration Maintenance Do not immerse The device is designed for use with Pall s Fluid Path only The unit is always on standby with battery power Avoid pressing po...

Page 9: ...Path Installation Loading indicator is illuminated Pump over temperature limit Check the temperature of the operating environment and allow pump to cool Pump slowing down with or without Battery discharging Continue filtration using power transformer plugged into main yellow fault indicator flashing power source until battery is fully charged Yellow fault indicator flashing Battery is discharged C...

Page 10: ...ltage 12 0 V Type NiMH nickel metal hydride DC Supply Voltage 24 0 V AC DC Power Transformer 24 VDC 1 6 Amp 100 240 VAC 47 63 Hz Input Interface IEC 320 3 pin socket Output Interface Power Entry Plug 2 5 mm barrel jack Power Cord IEC 320 connector and European CEE 7 7 plug IEC 320 connector and UK BS1363 plug IEC 320 connector and NEMA 5 15P grounded plug Disposal Instructions Reuse recycling and ...

Page 11: ...www pall com lab 11 ...

Page 12: ...ia Indonesia Ireland Italy Japan Korea Malaysia New Zealand Norway Philippines Poland Russia Singapore South Africa Spain Sweden Switzerland Taiwan Thailand United Kingdom and Vietnam Distributors in all major industrial areas of the world To locate the Pall office or distributor nearest you visit www pall com contact The information provided in this literature was reviewed for accuracy at the tim...