Figure 8
One way valve
Reservoir lid
Diafiltration
vessel
Female luer – barb
(to filtrate port)
Filtrate
collection
vessel
14. Connect “short tube” to fitting on front of gauge
mount and to 3-way valve on reservoir inlet. Attach
adjustable screw clamp to tube. Remove clamp
when not in use, tubing can deform over time
without fluid passing through tubing. (See figure 7)
15. Cut a length of tubing from Minimate Capsule
packaging long enough to reach filtrate collection
vessel from filtrate port on filter capsule. Attach
tubing clamp and female luer-1/8” hose barb to tube
Connect fitting to filtrate port and place open tube
end into filtrate receiver vessel. (See figure 8)
16. If performing diafiltration, place lid onto reservoir.
Connect 1-way valves onto fittings on top of
reservoir. From the Minimate Capsule filter
packaging, cut a length of tubing to reach from
1-way valve on reservoir lid top to diafiltration vessel
Attach male luer-1/8” hose barb elbow and tubing
clamp. Attach elbow to 1-way valve on lid and other
tube end into diafiltration vessel, sinker at end of
tube optional. (See figure 8)
Operation
■
Confirm all fittings are secure prior to running liquid
through system
■
Refer to VWR stir plate operating manual for
operation of stir plate. Use of reverse mode is not
recommended on stir plate with Minimate EVO
TFF System. If stir bar stops rotating when increasing
stir plate speed, reduce speed until stir bar is
again spinning
■
Refer to MasterFlex Pump and Pump Head
manuals for use
■
Refer to Minimate Instructions for use for operation
of the TFF capsule
Figure 7
Retentate port
Screw
clamp
Short
tube
Mid
length
tube
Male luer coupler
(to feed port)