Page 1: ...1 Athalon CM Crixus User Manual ...

Page 2: ...n 1 2 System Components 1 2 1 WiFi Gateway 1 2 2 Fluid Interface 1 2 3 Sensor Manifold 1 2 4 Filter Node 2 Installation Setup 2 1 Filter Node Installation 2 2 Providing a WiFi Access Point 2 3 Configuring the Athalon CM Gateway 3 Operation 4 Crixus Web Page 5 Mobile App 6 Troubleshooting ...

Page 3: ... sensors The modular design allows the end user maximum flexibility to efficiently integrate Athalon CM into their equipment arrangement without the expense of redundant components Minimum Athalon CM System includes WiFi Gateway SKU 1 Fluid Interface SKU 2 with optional Sensor Manifold Filter Node High pressure Filter Node SKU 3 4 or Return Line Filter Node SKU 5 ...

Page 4: ...orming utility connections and or installations 1 2 Pall System Components Pall Corporation is supplying a system comprised of WiFi Gateways Filter Nodes and Fluid Interfaces to monitor the health of the hydraulic fluid within the system Filter Nodes and Fluid Interfaces monitor the state of the fluid directly through attached sensors and report wirelessly to the Wi Fi Gateway The Gateway then rel...

Page 5: ...ccept any interference received including interference that may cause undesired operation The Wi Fi Gateway does not contain any user serviceable parts Changes or modifications not expressly approved by Pall Corporation could void the user s authority to operate the equipment Note This equipment has been tested and found to comply with the limits for a Class A digital device pursuant to part 15 of...

Page 6: ...erférence et 2 cet appareil doit accepter toute interférence notamment les interférences qui peuvent affecter son fonctionnement 1 2 1 4 Power Connection CAUTION To ensure the safe operation of the equipment the WiFi Gateway must be powered from an NEC Class 2 power supply that is certified as a Limited Power Source LPS Pall Corporation requires that certified electricians complete all wiring conn...

Page 7: ...end user piping systems 1 2 2 2 FCC Statement This device complies with Part 15 of the FCC Rules Operation is subject to the following two conditions 1 this device may not cause harmful interference and 2 his device must accept any interference received including interference that may cause undesired operation The Fluid Interface does not contain any user serviceable parts Changes or modifications...

Page 8: ...The device complies with Industry Canada s license exempt RSSs Operation is subject to the following two conditions 1 this device may not cause harmful interference and 2 this device must accept any interference received including interference that may cause undesired operation Cet appareil est conforme aux normes d exemption de licence RSS d Industry Canada Son fonctionnement est soumis aux deux ...

Page 9: ...ensor Node DC power connector must comply with the following requirements 1 Connector on Sensor Node end of power cable must be a 4 socket M8 style connector that will mate with Phoenix Contact 1453481 2 Cabling must comply with Article 725 of the NEC for Class 2 circuits 1 2 2 5 Power Supply Installation TBD 1 2 2 6 Sensor Connections The Fluid Interface provides two ports for connection to exter...

Page 10: ... FPS2800B12C4 1 2 2 8 Water Sensor FPS Sensor Fluid Viscosity Density Temperature TAN Seal Material Viton Fluorocarbon Housing Material Stainless Steel Port Option M14 x 1 5 mm male To fit Port 2 Manifold Maximum Working Pressure 20 bar 290 psi Rated Fatigue Pressure 20 bar 290 psi Measurement Viscosity 0 5 to 50 0 cp Measurement Density 0 65 to 1 50 gm cc Measurement Dielectric Constant Qualitati...

Page 11: ...Water Sensor WS12 Water in oil sensor Water content in oil RH or PPM and Temperature C or F Seal Material Viton Fluorocarbon Housing Material Stainless Steel Port Options NPT Internal Pipe Thread to ANSI B2 1 BSP G to ISO228 To Fit Port 3 Manifold Maximum Working Pressure 20 bar 290 psi Rated Fatigue Pressure 20 bar 290 psi Process Fluid Temperature 40 90 C 40 194 F Ambient Temperature 40 60 C 40 ...

Page 12: ...ISO228 Maximum Working Pressure 20 bar 290 psi Rated Fatigue Pressure 20 bar 290 psi Seal Material Fluorocarbon Manifold Material SG Iron Chemical Black Sensor Port 1 Options Internal O RING Boss Per SAE J514 BSP G to ISO228 Sensor Port 2 M14 x 1 5mm Internal O RING Boss Per ISO 6149 1 Sensor Port 3 Options NPT Internal Pipe Thread to ANSI B2 1 BSP G to ISO228 Physical Dimensions Diameter 120mm x ...

Page 13: ...l environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense This device u...

Page 14: ...DC Output Voltage Range 24 VDC 10 2 DC Output Current 1 0 A Minimum 3 Supply must be an NEC Class II power supply that is certified as a Limited Power Source LPS The cabling used to connect the external DC supply to the Filter Node DC power connector must comply with the following requirements 1 Connector on Filter Node end of power cable must be a 4 socket M8 style connector that will mate with P...

Page 15: ...Housing Material Brass Port Options 16 UNF Thread To BS1580 To Fit housing Pall DPI Port Maximum Working Pressure 450 bar 6500 psi Proof Pressure 675 bar 9800 psi Burst Pressure Typical 1100 bar 15900 psi Rated Fatigue Pressure 0 400 0 bar 5800 psi Process Fluid Temperature 25 90 C 13 194 F Ambient Temperature 40 60 C 40 140 F Measurement Differential Pressure Option 0 25 1 0 bar 3 6 14 5 psi 1 0 ...

Page 16: ...0 F Maximum Working Pressure 400 bar 6000 psi Rated Fatigue Pressure 240 bar 3500 psi Filter Element P N and Grade Read from RFID tag on Filter Element Filter Element Runtime Data Stored on RFID Tag Filter Element Lot Number Data Stored on RFID Tag Filter Element Activation Date Data Stored on RFID Tag Filter Element Hours in Service Data Stored on RFID Tag Filter Element Machine ID Data Stored on...

Page 17: ...Activation Date Data Stored on RFID Tag Filter Element Hours in Service Data Stored on RFID Tag Filter Element Machine ID Data Stored on RFID Tag Filter Element dP Max Data Stored on RFID Tag Filter Element Temperature Max Data Stored on RFID Tag Power requirement 24VDC 100mA Ingress Rating Electronics IP65 Minimum Compatibility Compatible with all petroleum oils and synthetic hydraulic and lubric...

Page 18: ...PA AES TKIP WPA2 AES TKIP Mixed mode Each gateway comes with a unique MAC address assigned by the factory which may also be used for authentication Upon initial power up the WiFi Gateway will function as a WiFi Access Point Using a laptop or similar device connect to the gateway s network SSID which is set to its MAC address Once connected browse to the Configuration page at 192 168 10 1 7000 Conf...