3
Calibration
12
3.3.2 0% Calibration
1
Rinse the probe well with DI water
or rinse solution.
2
Press
MODE
to select % saturation.
3
Dip the probe into zero oxygen solution.
Wait for the reading to stabilise.
4
Press
CAL
. The CAL indicator and intended
percentage calibration point (0%) display
briefly, before CAL flashes. The % value
based on the factory default calibration
is shown.
5
Press
HOLD/ENTER
to confirm the
calibration. The meter displays “CO”,
automatically calibrates to 0.0% saturation,
then returns to measurement mode.
Micro600DO_ZI INST 1240 16/9/15 12:57 Page 12