palintest DigiPAsS Operation Manual Download Page 2

Stage 3: Sample Preparation

Stage 4: Arsenic Measurement

Stage 5: Using the DigiPAsS

Take 

fl

ask and 

fi

ll to the 50ml line

with sample water and place on

fi

rm a surface.

 

Immediately push down bung device 

fi

rmly into the 

fl

ask.

Wait approximately 20 minutes (refer to timer on the  DigiPAsS).

After 20 minutes remove black slide and proceed to stage 4.

Remove Black Filter Slide

from Bung device. 

If visual reading is LESS than

100µg/l ppb). Insert Black Arsenic

Filter Slide into DigiPAsS.

Observe and Record result in µg/l

(micrograms per litre). If preferred 

record in (ppb) i.e 38µg/l = 38 ppb.

Finally open both Filter Slides and remove Filter Paper

from slide holders and dispose of in waste bag provided.

Record corresponding Colour. Match from colour chart

to obtain result in µg/l (ppb). 

If test result reads above 500µg/l (ppb) use dilution 

tube with de-ionised water and convert accordingly.

Then go back to stage 1 and do test again.

Black Filter Slide will

collect the Arsenic gas.

Red Filter Slide removes

excess Arsine gas.

For running concurrent multiple

tests use DigiPAsS Multipack

(X5 

fl

asks & X5 bung devices).  

Take A1 powder sachet and 

pour into the 

fl

ask.

 

Pick up loaded bung device and check 

fi

lter slides are fully pushed into bung.

Take tablet A2 from container and drop

carefully into the 

fl

ask, sample will start to 

fi

zz.

 

50ml

A

1

A2

#

*

X5

X2

X3
X4

X1

#

*

µg/l