palintest DigiPAsS Operation Manual Download Page 1

 

 

   

          

Graduated

Flask

Take one Hydrogen Sulphide Removal Filter

and insert into the bottom end of the bung.

 

Take Red Filter (Removal) Slide

and open.

 

Take the DigiPAsS and press any key to switch on.

This will then say 'Insert Slide'.

 

Take the pre-loaded Black Arsenic Filter Slide 

(the one that has just been 

fi

tted with 

fi

lter paper) and insert slide.

 

Then DigiPAsS will instruct to 'Remove Slide', once slide is removed this

then instructs to 'Run Test’, which automatically activates timer on the DigiPAsS. 

Then take the Black Filter Slide and insert 

into the bottom slot of the bung.

Take one Filter Paper from red labelled pot, use forceps and/or

gloves to insert into position. (repeat 

fi

lter paper insertion as before) 

 

 

Get Bung device and ONLY insert 

the Red Filter (Removal) Slide into the top slot.

Take Black Arsenic Filter Slide and open.

 

Take one Filter Paper from black labelled pot, use

forceps and/or gloves to insert into position ensuring 

the 

fi

lter paper is evenly covering the hole.Then 

fi

rmly close slide holder, taking care to locate the pins 

in the slide holder correctly.

Dilution Tube

Cleaning

Brush

Colour Chart

Waste Disposal Bag,

Gloves & Forceps

Tri-Filter Arsenic Trap

(Bung)

Operation Equipment Key

Stage 1: Loading bung device

Stage 2: Calibrating the DigiPAsS

Hydrogen Sulphide

Removal Filters

Black Filter Slide

(Detection)

DigiPAsS

Red Filter Slide

(Removal)

Screwdriver

DigiPAsS Operation Manual 

Filter Paper (arsenic test)

(Labelled Black)

Filter Paper (removal)

(Labelled Red)

Tablets A2

Powder Sachet

#

*

A 1

A2

#

*

*

#

#

*

#

*

Stage 3: Sample Preparation

Stage 4: Arsenic Measurement

Stage 5: Using the DigiPAsS

Take 

fl

ask and 

fi

ll to the 50ml line

with sample water and place on

fi

rm a surface.

 

Immediately push down bung device 

fi

rmly into the 

fl

ask.

Wait approximately 20 minutes (refer to timer on the  DigiPAsS).

After 20 minutes remove black slide and proceed to stage 4.

Remove Black Filter Slide

from Bung device. 

If visual reading is LESS than

100µg/l ppb). Insert Black Arsenic

Filter Slide into Arsenator.

Observe and Record result in µg/l

(micrograms per litre). If preferred 

record in (ppb) i.e 38µg/l = 38 ppb.

Finally open both Filter Slides and remove Filter Paper

from slide holders and dispose of in waste bag provided.

Record corresponding Colour. Match from colour chart

to obtain result in µg/l (ppb). 

If test result reads above 500µg/l (ppb) use dilution 

tube with de-ionised water and convert accordingly.

Then go back to stage 1 and do test again.

Black Filter Slide will

collect the Arsenic gas.

Red Filter Slide removes

excess Arsine gas.

For running concurrent multiple

tests use DigiPAsS Multipack

(X5 

fl

asks & X5 bung devices).  

Take A1 powder sachet and 

pour into the 

fl

ask.

 

Pick up loaded bung device and check 

fi

lter slides are fully pushed into bung.

Take tablet A2 from container and drop

carefully into the 

fl

ask, sample will start to 

fi

zz.

 

50ml

A

1

A2

#

*

X5

X2

X3
X4

X1

#

*

µg/l