7.
Enable the Output with no load connected
8.
Enter the DMM VAC readings for all three phase in the respective Voltage data entry
text boxes in the Calibration screen. Move the DMM probe from phase A, to B to C
respectively to obtain each phase reading.
9.
Press "Cal V. gain" soft key
10.
Wait for the V gain coefficients for all phases to be calculated and displayed
11.4.4
Current Gain Calibration Setup Diagrams
For current calibration, a resistive load equivalent to 90% of maximum available RMS phase
current for the phase mode selected is recommended. It is permissible to use only one load and
move it between phases for three-phase mode calibration. For single-phase mode current
calibration, a load capable of supporting 90% of maximum available single-phase current is
recommended. The A, B and C outputs must be shorted together for single-phase mode
operation. The optional Single Phase Shorting Connecter (P/N 160086) can be used to
accomplish this.
Note:
To guarantee that the current doesn’t change during the calibration process, it is
recommended that the unit is operated at current limit, but reducing its set point to the
desired value. It is also recommended to use a stable load that provides a constant
linear impedance to the unit.
A suitable current shunt must be connected in series with the load as shown below for three-
phase mode. The sense terminals of the current shunt must be connected to the DMM Voltage
input.
Three Phase Mode Setup
AC Line
Input
DMM
Hi Sense
Lo Sense
R Load
A B C N N N
Figure 11-2: Current Calibration Equipment Setup – 3 Phase Mode – Phase A