DOO
A-Amp, V-Amp , AVI
V-sense, A-sense,Aref
DVI
A-Amp, V-Amp, AVI, V-sense
A-sense
DDD
A-Amp, V-Amp, AVI, A-sense, V-
sense,Aref
-----
VDD
V-Amp, A-sense, V-sense, Aref
A-Amp, AVI,
VVI
V-Amp, V-sense
A-Amp, A-sense, AVI,Aref
VOO
V-Amp
A-Amp, V-sense, A-sense, AVI,
Aref
6.6
Demand and asynchronous Modes
Demand Modes available are AAI > DVI > DDD > VDD > VVI
Asynchronous Modes available AOO > DOO > VOO
Knobs turned clockwise select Mode starting from
AOO > AAI > DOO > DVI > DDD > VDD > VVI > VOO>Urg
6.7
Urgent Mode
For urgent mode setting turn mode selection knob clockwise to
extreme right.
6.8
Collet terminals
Accepts External-pacing leads ranging from 1.8 to 2.3 mm diameter.
6.9
Lead placement
a). When positioning the lead in the heart it is recommended that the
ECG should be monitored in order that the S-T segment elevation,
indicating contact with the myocardium and impaction, can be noted.
b). Connecting the lead to the ZEUS-1
Switch OFF the device (Zeus-1).
Connect the proximal electrode to the indifferent (+) red terminal and
the distal electrode to the active (-) black terminal of Temporary lead