REPAIR
POWER SOURCE
Contact PACE Customer Service at Tel. (301) 490-9860, FAX (301) 604-9215 to obtain any replacement parts.
Refer to the "Replacement Parts" section of this manual for part numbers.
Table III. Corrective Maintenance, Power Source
S Y M P T O M
PROBABLE CAUSE
S O L U T I O N
Digital Readout is blank.
No functions on system.
Blown Fuse (F1)
Replace Fuse F1 located on rear of
Power Source in the AC Receptacle.
E-1 displayed on Digital
Readout.
All Channels are unplugged
on Thermal Mgmt. Center.
Plug SensaTemp handpiece into CH 1,
CH 2 OR CH 3.
Open sensor in SensaTemp
handpiece.
Refer to handpiece Operation Manual for
Corrective Maintenance procedures.
E-2, E-3 or E-4 displayed
on Digital Readout.
Defective heater assembly in
SensaTemp handpiece.
Unplug all SensaTemp handpieces.
Plug handpieces back in one at a time.
When defective handpiece is
connected, "E" code will be displayed.
Refer to handpiece Operation Manual.
No heat on SensaTemp
handpiece(s) at Thermal
Management Center.
Digital Readout displays
very low temperature.
Open heater.
Refer to handpiece Operation Manual.
Blown 12 Amp Fuse.
Replace 12 amp Fuse located on rear of
Power Source.Disconnect all
SensaTemp handpieces before
powering system up. Plug SensaTemp
handpieces back in one at a time. If the
fuse blows when a handpiece is plugged
in, the handpiece is shorted. Repair the
handpiece.
Handpieces connected to
AUX receptacles.
SensaTemp handpieces are
not AUX compatible.
Disconnect incompatible handpieces
from AUX Receptacles. Connect to
Power Receptacles on front panel.
Damaged Power Receptacle. Replace Power Receptacle.
System overload. Check for
defective SensaTemp
handpiece.
Disconnect defective handpiece.
220 Volt version only: Overtemp cutout
in transformer may open. Allow
transformer to cool.
Microprocessor pcb defect.
Replace Microprocessor pcb.
Summary of Contents for PRC 2000
Page 1: ...SERVICE MANUAL PRC 2000 SYSTEMS ...
Page 2: ...i MANUAL NO 5050 0344 REV C ...
Page 14: ...9 GENERAL INFORMATION Figure 5 Pik And Paste Section ...
Page 18: ...13 GENERAL INFORMATION ...
Page 20: ...15 GENERAL INFORMATION Figure 7 Rear Panel Parts I D ...
Page 28: ...23 REPAIR Figure 9 Connector Plug Wiring Figure 10 MicroChine Wiring MICROCHINE HANDPIECE ...
Page 43: ...38 REPAIR ...
Page 45: ...40 REPAIR POWER Figure 16 Power Flow Chart ...
Page 46: ...41 REPAIR ...
Page 47: ...42 REPAIR TMC Figure 17 Thermal Management Center Malfunction Flow Chart I ...
Page 48: ...43 REPAIR ...
Page 49: ...44 REPAIR Figure 18 Pik Paste Malfunction Flow Chart PIK PASTE ...
Page 50: ...45 REPAIR ...
Page 51: ...46 REPAIR Figure 19 MicroChine Malfunction Flow Chart MICROCHINE ...
Page 52: ...47 REPAIR ...
Page 53: ...48 REPAIR PULSE PLATE Figure 20 Pulse Plate Malfunction Flow Chart ...
Page 54: ...49 REPAIR ...
Page 55: ...50 REPAIR PULSE HEAT Figure 21 Pulse Heat Malfunction Flow Chart ...
Page 56: ...51 REPAIR ...
Page 57: ...52 REPAIR Figure 22 PPS 400 PPS 400J PPS 400E Wiring Diagram WIRING DIAGRAM ...
Page 58: ...53 REPAIR ...
Page 60: ...55 REPAIR ...
Page 62: ...57 REPAIR ...
Page 64: ...59 REPAIR ...
Page 66: ...61 REPAIR ...
Page 68: ...63 REPAIR ...
Page 69: ...64 REPAIR Figure 25 Display PCB Assembly Schematic DISPLAY PCB ASSEMBLY SCHEMATIC ...
Page 70: ...65 REPAIR ...
Page 71: ...66 REPAIR ASSEMBLY DRAWING Figure 26 Assembly Drawing ...
Page 72: ...67 REPAIR Figure 27 Assembly Drawing Cont d ...
Page 73: ...68 REPAIR AIR HOSE ROUTING Figure 28 Air Hose Routing ...
Page 75: ...70 REPLACEMENT PARTS Figure 29 Power Source Replacement Parts POWER SOURCE CONT D ...
Page 76: ...71 REPLACEMENT PARTS Figure 30 Power Source Replacement Parts Cont d ...
Page 77: ...72 REPLACEMENT PARTS Figure 31 Power Source Replacement Parts Cont d ...