INTRODUCTION 

 

The following procedure will allow you to effectively calibrate your MET 250/220 system to insure 
accuracy and maintain peak performance. 

Perform the following procedure step by step, in sequence, to: 
 

1.  Alter Tip Temperature values. 
2. Alter 

Tip Temperature Offset values. 

3.  Determine and/or alter feature status and defaults stored in the system. 
4.  Recalibrate the system for accuracy. 

 

PROCEDURE  

KEYLOCK OPTION 

 

KEYLOCK 
1. An optional KeyLock feature is available from PACE which 
prevents unauthorized alteration of stored data or calibration of the 
system. Check the rear panel of the system power source. If the 
KeyLock feature is present there will be a KeyLock switch located in 
the upper left portion of the panel. Use the key to turn the switch to 
the "Unlock" position. If the feature is not present, there will be a 
round plastic filler plug present at that location.

 

NOTE 

 

The KeyLock switch-must be turned  to the "UNLOCK!" position to alter 
any of the data stored in the memory or to recalibrate the system

 

Summary of Contents for MBT 220

Page 1: ...or alter feature status and defaults stored in the system 4 Recalibrate the system for accuracy PROCEDURE KEYLOCK OPTION KEYLOCK 1 An optional KeyLock feature is available from PACE which prevents una...

Page 2: ...ld the Tip Temperature Set and Scroll Down Keys 4 Place POWER Switch in the ON I Position All of the system LEDs will light The Temperature Display will read 888 and change to read 1 X 5 Release the T...

Page 3: ...efault AUTOMATIC SETBACK 8 As received from the factory the temperature will read S meaning that Automatic Setback is turned off A readout of 1 thru 9 on the right side of the Digital Readout indicate...

Page 4: ...are necessary 11 The Channel 1 LED is now lit signifying that Channel 1 is ready for calibration Perform steps 12 through 21 to calibrate As channels change repeat these steps for each channel NOTE Al...

Page 5: ...oll Up Key or decrease Scroll Down Key the Lower Temperature Limit value 14 Press and release the Tip Temperature Set Key to store the display value into memory 15 The Digital Readout now displays H X...

Page 6: ...ou do not wish to recalibrate for readout accuracy press the Tip Temperature Offset Key and perform steps 12 through 17 After all channels have been calibrated you may exit the Calibration Mode by pre...

Page 7: ...microprocessor controlled circuitry is recalibrating another aspect of the system circuitry CAL will now be displayed indicating that calibration of this channel is complete 22 Press and release the T...

Page 8: ...ghts Reserved Printed in the USA www paceworldwide com TEMPERATURE DISPLAY MESSAGE CODES Listed below are message codes and a description of each which may be displayed on the system power source digi...

Page 9: ...MAKES NO OTHER WARRANTY EXPRESS OR IMPLIED AND MAKES NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE Seller will at its option repair or replace any defective products at its facil...

Reviews: