I
NSTRUCTION
M
ANUAL
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1 520-882-6598 Fax +1 520-882-6599 email: [email protected] Web: http://www.metallographic.com
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
OMEGA-RT-60/150 Rockwell Hardness Tester
23. Connecting rod
24. Adjustment screw
25. Fine tuning adjustment
knob
2.3.7 Calibration
To calibrate to test blocks -
hold screw in place at (24)
and loosen nut, adjust
screw (25)
Values reading low - adjust
screw clockwise
Values reading high - adjust
screw counter-clockwise
10