3.4 Safety Tests
!
Examine and verify that the
MEGA-M300
safety
devices and operating performance are in good working
condition prior to use. The following safety checks are
considered important:
Emergency stop switch
Magnetic Safety switch
9
Test:
Activate main switch and close hood;
depress emergency stop switch.
Proper
Response:
Machine powers down.
Malfunction:
Machine does not lose power.
Corrective
measure:
If system does not power down,
disconnect power supply cord and call
service technician.
Magnetic safety lock out switch
Test:
Activate main switch and close hood;
turn cut-off wheel ON then OFF;
Try to open the hood.
Proper
Response:
Hood does not open for approximately 5
seconds.
Malfunction:
Hood opens.
Corrective
measure:
If cut-off wheel does not power down,
disconnect power supply cord and call
service technician.
Emergency stop switch