34
Notes:
•
Only products that have been effectively cleaned and disinfected are allowed to be sterilized.
•
Before using the sterilizer for sterilization, read the Instruction Manual provided by the
equipment manufacturer and follow the instructions.
•
Do not use hot air sterilization and radiation sterilization as this may result in damage to the
product.
•
Please use the recommended sterilization procedures for sterilization. It is not recommended
to sterilize with other sterilization procedures such as ethylene oxide, formaldehyde and low
temperature plasma sterilization. The manufacturer assumes no responsibility for the
procedures that have not been recommended. If you use the sterilization procedures that
have not been recommended, please adhere to related effective standards and verify the
suitability and effectiveness.
* Fractional pre-vacuum procedure = steam sterilization with repetitive pre-vacuum. The
procedure used here is to perform steam sterilization through three pre-vacuums.
6.4.9 Storage
1.
Store in a clean, dry, ventilated, non-corrosive atmosphere with a relative humidity of 10% to
93%, an atmospheric pressure of 70KPa to 106KPa, and a temperature of -20°C to +55°C
(-4°F to
131°F).
2.
After sterilization, the product should be packaged in a medical sterilization bag or a clean
sealing container, and stored in a special storage cabinet. The storage time should not
exceed 7
days. If it is exceeded, it should be reprocessed before use.
Notes:
•
The storage environment should be clean and must be disinfected regularly.
•
Product storage must be batched and marked and recorded.