background image

MAINTENANCE

DESCRIPTION OF SYMBOLS

WARNING! 

BEFORE CLEANING YOUR TOOL, MAKE 

SURE THAT THE POWER SUPPLY HAS BEEN 

REMOVED OR DISCONNECTED TO PREVENT 

ACCIDENTAL STARTING.

Warning

Regulatory Compliance Mark

(RCM)

V

Volts

%RH

°F

Percent Relative Humidity

°C

Degrees celsius

Degrees Fahrenheit

Read Instruction Manual

Power tools that are no longer usable should not be 

disposed of with household waste but in an environmentally 

friendly way. Please recycle where facilities exist. Check 

with your local council authority for recycling advice.

Recycling packaging reduces the need for landfill and raw 

materials. Reuse of recycled material decreases pollution in 

the environment. Please recycle packaging where facilities 

exist. Check with your local council authority for recycling 

advice.

CARING FOR THE ENVIRONMENT

SPARE PARTS

Spare parts can be ordered from the Special Orders Desk 

at your local Bunnings Warehouse.

 

For further information, or any parts not listed here, visit 
www.ozito.com.au or contact Ozito Customer Service: 
Australia 1800 069 486 
New Zealand 0508 069 486 

E-mail: [email protected]

Cleaning

Clean the appliance regularly with a soft, dry cloth. Do not use 

cleaning agents or solvents; these may be aggressive to the plastic 

parts in the appliance. Ensure that no water can get into the interior of 

the appliance.

Calibration

A standard sound source which outputs 94dB, 1kHz sine waves 

should be used to calibrate the unit. You will need a calibrator (not 

supplied) to perform this task. It is recommended that calibration of 

the meter is carried out once a year.

1.  Remove the windshield if it is 

attached.

2.  Press and hold down the ‘FAST | 

SLOW’ and ‘MAX | MIN’ buttons.

3.  Then press the power button until the 

backlight flashes briefly.

Note:

 ‘CALI’ will appear on screen once 

the meter has entered the calibration 

mode.

4.  Follow the manufacturer’s instructions for using your calibrator.

5.  Press the ‘BACKLIGHT | HOLD’ 

button to start calibrating the unit. The 

display will start flashing.

6.  After 2 seconds, calibration is 

complete and the meter will switch off 

automatically.

Storage

Switch off the tool, place it in the storage bag and store it in a dry 

location which is not accessible to unauthorised persons.

Remove the batteries before storage if the tool will not be in use for a 

long time.