Oxford PRECISION WS-6E Instruction Manual Download Page 4

3. Working temperature: 

5

°

C-35

°

C.

4. Power supply:

12Vdc, 650mA AC adaptor, with negative centre or 6V4AH rechargeable

battery.

Note:

• When the rechargeable battery is fully charged the scales can be used for 20 hours

continuously.

• When the power supply registers <5.4V(±0.1V), the scales will be turned off

automatically.

• When the power supply registers <5.7V(±0.1V) but over 5.4V(±0.1V), the scales will

alternatively display 

Lo.bAT

and whatever the screen currently displays in its normal

mode.

• If the AC adaptor is used to charge the battery or to power the scales, the switch on

the underside of the scales must be switched to the 

on

position.

5. Working current:

<150mA.

6. Recharging current:

<400mA.

1. ZERO/ON/OFF

(i) Press this button to activate the scales. The scales perform its self-examination mode,

display the working voltage 

x.x Udc

, display the machine’s capacity (F6.000kg or

F15.000kg), before finally setting the zero point and displaying a zero reading.

Note: The power-on zero-point range is: calibration zero-point ±20% FS (full scale weight

capability).

(ii) To reset the scales to zero after any operation, press the 

ZERO

key for less than five

seconds. Pressing the 

ZERO

key for more than five seconds will activate the 

ON/OFF

function.

The ZERO key range is power-on zero-point ±5% FS. The

ZERO

function is used to clear

any 

TARE

weight that may be displayed after previous weighing operations. When the

TARE

weight is displayed the 

NET LED

will be activated. After using the 

ZERO

function

the 

ZERO LED

will be shown. If the current weight exceeds ±5% then the 

ZERO

operation is disabled and the display will read 

0

______

or 

0______

, before returning to

the previous reading.

2. TARE/UNIT

Pressing the 

TARE/UNIT

key for less than five seconds activates the 

TARE

function.

Pressing the 

TARE/UNIT

key for more than five seconds activates the 

UNIT

function.

The 

TARE

range for these scales is up to 102%FS. The 

TARE

function should be

activated when the scales are in normal mode, i.e. the display reads 0 and no items are

on the weighing tray.

When the 

TARE

weight is not reading 0, the NET LED light will be activated.

Note: Gross weight = Net  Tare weight 

£

102%FS.

UNIT:

The 

UNIT

function is used to select the weight scale (lb, kg, g, oz, lb.oz).

Key Functions

Calibration

AUTO-OFF function

Warnings

Cleaning the scales

The calibration of this product requires specialist tools and should only be carried out by

persons with appropriate skills/qualifications. Should you experience any problems please

return the goods to your supplier, who will arrange professional calibration of the scales.

UNDER NO CIRCUMSTANCES SHOULD MAINTENANCE OR CALIBRATION BE ATTEMPTED

BY PERSONS WITHOUT APPROPRIATE SKILLS OR QUALIFICATIONS. SUCH ACTIONS MAY

RESULT IN IRREPAIRABLE DAMAGE AND WILL ALSO INVALIDATE ANY GUARANTEE OR

WARRANTY.

The scales incorporate an 

AUTO-OFF

function. To activate this action the scales must be

switched on. To set the function, press the 

TARE

and 

ZERO

buttons for more than 5

seconds until the 

A.oFF.x

message is displayed.

Setting the 

AUTO-OFF

function time is achieved by pressing the 

TARE

button the required

number of times, until the desired 

AUTO-OFF

time is displayed (i.e. 

A.oFF.4

will cause the

scales to switch themselves off after four minutes).
The 

AUTO-OFF

time is confirmed and saved by pressing the 

ZERO

button.

If the scales are to be used in wet or damp conditions using a mains electrical connection

(110V/230V), the supply must be correctly fused and protected by an approved RCD which

conforms to BS7071 or BS7288 or IEC1008-2-2 BS EN 61008:1995. 

IF YOU ARE NOT

SURE ABOUT THIS WARNING YOU MUST SEEK THE ADVICE OF A QUALIFIED

ELECTRICIAN.

To avoid an electric shock, or permanent damage to the machine, you must isolate the

scales from the electricity supply before carrying out any routine maintenance.

To avoid damage to the display, top plate or body of the scales, it is not advised to use any

harsh abrasives, scouring cleaners and alkaline solutions, such as washing soda, when

cleaning the scales. Special care should be taken when cleaning the digital display screen,

which could be permanently damaged by the use of inappropriate cleaning agents. 

DO NOT

ATTEMPT TO CLEAN ANY INTERNAL COMPONENTS.

Best cleaning results are obtained by the use of a clean, lint-free cloth, which has been

slightly dampened using a weak water and washing-up liquid solution.

Summary of Contents for WS-6E

Page 1: ...c or imperial Capacities WS 6E 6kg 12lb 12oz WS 15E 15kg 30lb Oxford Precision Victoria Works Leicester England OXFORD PRECISION PRECISION Quality Guarantee Warranty Quality Guarantee Oxford Precision products carry a one year manufacturers warranty Oxford Precision products are designed and manufactured to the highest standards and specifications Assuring the quality and performance required by a...

Page 2: ... be placed carefully onto the scales and the display allowed to stabilise Remove the item immediately Using the scales in this manner will increase the longevity and maintain the accuracy of the scales Using the scales in the vicinity of in use cell phones radios computers or other electronic devices which emit radio frequency energy Exposure to such emissions may cause the scales to malfunction a...

Page 3: ...ght exceeds power on zero point 5 FS when the ZERO button is pressed 0______ The power on zero point is below the calibration zero point 20 FS when the scale is turned on or the weight is power on zero point 5 FS when pressing ZERO button ______ The current weight is less than 25 FS ______ The current weight is over 102 FS CAL 0 Zero point at calibration CAL F Full capacity at calibration CAL Er C...

Page 4: ... will be activated Note Gross weight Net weight Tare weight 102 FS UNIT The UNIT function is used to select the weight scale lb kg g oz lb oz Key Functions Calibration AUTO OFF function Warnings Cleaning the scales The calibration of this product requires specialist tools and should only be carried out by persons with appropriate skills qualifications Should you experience any problems please retu...

Reviews: