PROBE-RELATED CAUTIONS:
CAUTION
OxyFlo Pro probes must be HANDLED WITH CARE.
Failure to do so may result in breakage of the internal optical
fibres, scratching the polished probe ends or separation of
the cable from the probe ends or connectors.
CAUTION
The optical fibres used in OxyFlo Pro probes consist of glass
with a diameter of 125 µm. The fibres are flexible and can be
bent, however it is recommended that they are not subjected
to bends with a radius of less than 30mm.
CAUTION
DO NOT drop, apply tension or ‘kink’ any part of an OxyFlo
Pro probe. Permanent damage may result.
CAUTION
OxyFlo Pro probes are NOT approved for use on patients.
CAUTION
Do not use a worn or damaged probe.
CAUTION
OxyFlo Pro probes should be stored, with the probe cable
carefully coiled to avoid ‘kinks’, in the dedicated protective
case in which they are supplied.
CAUTION
Avoid using OxyFlo Pro probes under strong (e.g. surgical)
lights.