![Oxford Instruments OpAL System Manual Download Page 304](http://html.mh-extra.com/html/oxford-instruments/opal/opal_system-manual_1664943304.webp)
^ää=lfmq=póëíÉãë
lñÑçêÇ=fåëíêìãÉåíë=mä~ëã~=qÉÅÜåçäçÖó==
System Manual
OKQ
j~íÅÜáåÖ=ÅçãéçåÉåí=ä~óçìíë=
The layout of the matching components depends on the device to be matched to the RF
Generator to ensure maximum power transfer. Typical layouts of the components are shown
in Fig 3.
Note that in the typical layout, padding capacitors can be added in parallel with C1 and C2 to
modify their capacitance ranges. Refer to sub-section 5.3 (page 18) for details.
RF IN
RF OUT
DC BIAS
COIL
C1
COIL
C2
C2 PADDING
CAPACITORS
C1 PADDING
CAPACITORS
TO DC BIAS
CIRCUIT
TYPICAL MATCHING COMPONENTS
(LAYOUT 1)
RF IN
RF OUT
C1
C2
ALTERNATIVE MATCHING COMPONENT LAYOUT
(LAYOUT 2)
NOTE: In some systems the circuit labels C1 and C2 are reversed.
Fig 3: Matching component layouts
OIPT Automatch Unit
Issue 6: September 05
Page 6 of 20
Printed: 16-Nov-07, 9:26
Summary of Contents for OpAL