USER INTERFACE
Original Instructions
Gas Cfg
Tap once to access the Gas Flow Configuration page (see
PID Table
Tap once to access the PID Table Editor page (see
). This button is
only active if
PID
is set to
Auto
.
3.2.10
Gas Flow Configuration page
Figure 3-23 Gas Flow Configuration page
This page configures the gas-flow control loop, which is controlled by the optional
auxiliary I/O daughter-board. The page contains the following fields and controls:
Sensor
Select the sensor to be used for the gas-flow control from a drop-down menu.
TES
Enter the Temperature Error Sensitivity as a power of 2. e.g. enter 7 to set 512 (2
7
).
TES controls how fast the needle valve responds to a temperature error. Increasing
TES by 1 increases the response rate by a factor of 2.
TVES
Enter the Temperature Voltage Error Sensitivity as a power of 2. e.g. enter 6 to set
256 (2
6
). TVES controls how fast the needle valve responds to a heater-voltage error.
Increasing TVES by 1 increases the response rate by a factor of 2.
Min Gas Flow (%)
Enter a minimum valve position as a percentage of the maximum. Defining a
minimum valve position prevents the valve from closing fully, and possibly sticking.
Choose the minimum position so that there is no gas flow.
GFSF (%)
Displays the gas flow scaling factor.
HTT
Select a
Heater Target Table
file to use for the control.