Helium has been switched to FAST & current is on for a pulse.
X210S020000R00
Pulse is complete. Helium remains in FAST rate.
X210S0A0000R00 Still in FAST. Level has fallen below 90% (FULL threshold).
X210S0C0000R00 Operator has switched Helium back to SLOW.
X210S1A0000R10 Level has fallen below 20% (FILL threshold).
Filling has started and Relay 1 is active.
Rate has automatically switched to FAST.
X210S120000R10
Level has risen above 20%. Filling is continuing.
X210S020000R00
Level has risen above 90%. Filling has stopped.
X210S040000R00
Level has not risen for 15 minutes so rate switched to SLOW.
X210S0C0000R00 Level has fallen below 90%
X210S1A0000R10 Level has fallen below 20% (FILL threshold).
Filling has started and Relay 1 is active.
Rate has automatically switched to FAST.
X210S7A0000R96 Fill has failed. Level has fallen below 10% (LOW threshold).
ALARM lamp is lit, ALARM is sounding and Relay 4 is active.
X210S7A0000R14 Operator has pressed SILENCE button.
Sound is cancelled and Relay 4 has released.
Alarm lamp remains lit.
X210S1A0000R10 Filling has been resumed. Level is above 10%.
Alarm lamp goes out and filling continues.
X210S120000R10
Level has risen above 20%. Filling is continuing.
X210S020000R00
Level has risen above 90%. Filling has stopped.
X210S040000R00
Level has not risen for 15 minutes so rate switched to SLOW.