Cypher SPM
User Guide
28
28
Preparing For Conductive AFM On The
Environmental Scanner
Conductive AFM works only with the gas cell body, but with any sample stage.
Prerequisites: Assemble your sample stage and gas cell body, as described in
Mount the sample stage onto the environmental scanner, as described in
ging the sample stage and cell body on page 84
Install the gas cell cable, as described in
Gas cell cable installation on page 65
Follow the procedure
Zeroing the ORCA sample bias signal on page 190
Required
items:
An ORCA conductive AFM cantilever holder, such as 901.767 (see
tification of ES cantilever holders on page 54
).
The gas cell body 901.746 (see
).
The gas cell cable 448.137 (see
Gas cell parts list on page 63
).
The Electrical Sample Puck Assembly (448.140), included in the Environmental
Scanner Accessory kit.
An electrically conducting cantilever, such as the Asylum Electrilever.
1. Prepare the cantilever holder and AFM, as described in steps 1–17 of
sample on an environmental scanner on page 205
.
2. Remove old sample.
l
Use Platypus tweezers (part # 290.165) to remove the
old sample from the sample stage.
l
Wipe the sample stage clean with a cloth or use com-
pressed air.
Any dust or grit will prevent the sample disc from
seating properly.
28 Preparing for conductive AFM on the environmental scanner
Page 193 of 243