10.1.3 Gas cell cable installation
1. Raise the engage stage fully.
l
Rotate the Engage Control Knob on the Cypher clock-
wise and hold until the cantilever holder is far from
the sample or is at its upper limit of travel.
l
Although it is not required, for safety reasons we
recommend making motor moves with the door
closed. Beware of pinch points.
Note:
Pay attention! If you turn the knob the
wrong way (counterclockwise), you will lower the
cantilever holder instead of raising it. When you
lower the cantilever holder, you can crash the can-
tilever holder into the sample and cause serious
damage to the scanner.
2. Unlock the scanner and pull it all the way forward.
This provides sufficient access to plug in the cable.
Tip:
You might find it more comfortable to take the scanner out of the AFM completely. To
do so, turn off the system power and follow the instructions for
sample on an environmental scanner on page 205
.
3. Locate cable.
l
Locate the Gas Cell Cable (448.137).
4. Orient the cable.
l
The cable will plug in with the holes facing toward the
top of the scanner.