Cypher SPM
User Guide
20
20
Inserting The Cantilever Holder On A Standard
Scanner
Note:
If you own the environmental scanner, please follow the tutorial in
tilever holder on an environmental scanner on page 157
You need the following items to perform the tutorial:
Cantilever
holder
The cantilever holder with required cantilever installed for your experiment.
See
Changing the cantilever on a standard scanner on page 144
Tools
Part# 290.102: Tweezers with curved tips.
Part# 290.139: 1/16” ball head wrench.
1. Rotate the Engage Control knob clockwise until it reaches its upper limit of travel.
If you do not raise the stage, you will crash the cantilever into the sample and ruin the can-
tilever, sample, and possible the cantilever holder.
2. If you have not already done so, ensure that the scanner is already pulled forward and any exist-
ing cantilever holder is removed.
See
Removing the cantilever holder on a standard scanner on page 141
Page 148 of 243
20 Inserting the cantilever holder on a standard scanner