![Owon MSO5022 User Manual Download Page 102](http://html.mh-extra.com/html/owon/mso5022/mso5022_user-manual_1664706102.webp)
User’s Manual of
OWON
Color Mixed Signal Digital Storage Oscilloscope
Appendix A: Technical Specifications
Unless otherwise specified, the technical specifications applied are applicable to the
probe with the attenuation switch setting 10X and the MSO series digital oscilloscope.
Only if the oscilloscope fulfill the following two conditions at first, can these
specification standards be reached.
This instrument should run for more than 30 minutes continuously under the
specified operating temperature.
If the change range of the operating temperature is up to or exceeds 5
℃
, open the
system function menu and execute the “Auto-calibration” procedure.
All specification standards can be fulfilled, except one(s) marked with the word
“Typical”.
Digital Storage Oscilloscope
Sampling
ACQU Mode
Sample
Peak Detection
Average
Sampling Rate
100 MS/s
Input
Input Coupling
DC, AC
Input Impedance
1M
Ω
± 2%, connected with 20pF± 5pF in parallel
Probe attenuation coefficient 1X, 10X, 100X, 1000X
Max. Input Voltage
300V, Peak Value
Level
Sampling Rate Range
10S/s
~
100MS/s
Waveform interpolation
(sinx) /x
Record Length
5000 sampling points per channel
Scanning Speed Range
(S/div )
5ns/div
~
5s/div, according to the stepping mode of
1- 2.5- 5.
Accuracy of Sampling Rate
and Delay Time
± 100ppm ( for any time interval equal to or greater
than 1ms)
Measuring Accuracy of Time
Interval ( T)
△
(Full
Bandwidth)
Single: ± (One Sampling Interval Time +100ppm ×
R0.6ns)
>Averages of 16: ± (One Sampling Interval Time
+100ppm× R0.4ns)
102