5
.Menu Interpretation
53
5.12
【
Source
】
hardkey
Key access:
Press
Shift key, then
Source
hardkey (
6
in
numeric keypad
).
When the Source is turned on, an independent signal or a signal
with the same frequency of the current sweep signal will be output
from the GEN Output
50Ω terminal on the
top panel. Press the key
to access the submenu includes [Output On Off], [Source GEN CW
TG], [Output Level], [Output FREQ], [OF Step], [Network Meas
►
].
The source is turned off in the power-on and reset states.
5.12.1 [Output On Off]
Turn on
/
off the output of the GEN Output
50Ω terminal on the
top
panel.
5.12.2 [Source Gen CW TG]
Select source as CW (Continuous Wave) or TG (Tracking
Generator).
CW:
Output independent continuous wave signal. (CW output
function is only for specific models.)
TG:
RF output and spectrum reception are fully synchronized on
the frequency sweep, and the tracking source frequency can not
be set individually.
5.12.3 [Output Level]
Set the output power of CW or TG source.
5.12.4 [Output FREQ]
When the source is “CW”, set the output frequency of CW source.
5.12.5 [OF Step]
When the source is “CW”, set the output frequency step of CW
source.