6

 

Troubleshooting 

Problem 

Indication 

Possible Cause 

Action  

System 

Indicator: 

 

Solid amber 

Air Filter is exhausted / not 
present / not detected 

Check Air Filter Lifetime 
LEDs. Replace Air Filter if 
current Air Filter is 
exhausted / not present / 
not detected. 

The system has detected 
an internal error when flow 
is off or within the first ten 
seconds of starting 

Reset system by hard 
power off and on using IEC 
power inlet rocker switch 
and try to start pump 
again. If this does not clear 
the error, request service. 

System 

Indicator: 

 

Flashing red 

Audible 

Alarm 

Blockage of the Air Supply 
Tubing while the system is 
providing air 

Remove air supply from 
test subject if present. Turn 
off flow using on/off switch. 
If possible, remove 
blockage and reset system 
by hard power off and on 
using IEC power inlet 
rocker switch. If not, 
replace Air Supply Tubing. 

Removal of the Air Filter 
from the unit while the 
system is providing air 

Remove air supply from 
test subject if present. Turn 
off flow using on/off switch. 
Return the air filter to the 
unit and reset system by 
hard power off and on 
using IEC power inlet 
rocker switch. 

Summary of Contents for BREATH BIOPSY CASPER

Page 1: ...01 0931 01 0937 CASPER Portable Air Supply User Manual CC 950006 TM 4 ...

Page 2: ...edical Ltd 2019 95 0006 004 08 Oct 2019 Page 2 of 25 Symbol Key Manufacturer Reference number Serial number Date of manufacture Consult instructions for use CE mark DO NOT dispose of in household waste Use by date ...

Page 3: ... Owlstone Medical Ltd 2019 95 0006 004 08 Oct 2019 Page 3 of 25 Storage Temperature Limit Humidity Limit Fragile handle with care Do not use if package is damaged ...

Page 4: ...ct 2019 Page 4 of 25 Contents 1 General Information 5 2 Introduction to the CASPER Portable Air Supply 10 3 Setup of the CASPER Portable Air Supply 12 4 Operating Instructions 17 5 Indications 20 6 Troubleshooting 22 7 Servicing 25 8 Cleaning 25 ...

Page 5: ...le Air Supply is for professional research and development use only The CASPER Portable Air Supply is only to be used in the presence of a trained user who is capable of removing the ReCIVA Breath Sampler from the test subject It is the responsibility of the user to monitor the test subject s wellbeing at all times while using the CASPER Portable Air Supply The CASPER Portable Air Supply is not a ...

Page 6: ...ch as solvents and biologically hazardous materials Ensure that the system is correctly assembled before use Do not place any objects directly on the CASPER Portable Air Supply Do not block the inlet or outlet ports of the system nor seal any tubing during operation If there is any doubt as to the operation of the system such as a drop in the air flowrate it should be powered down and disconnected...

Page 7: ...emoved from the test subject It is the responsibility of the operator of the CASPER Portable Air Supply to assess the impact of air supply failure and provide adequate mitigations should this occur The CASPER Portable Air Supply should not be used at altitudes above 2km in subjects who have not acclimatised The CASPER Portable Air Supply should not be used for longer than 1 hour by a single test s...

Page 8: ...ary to comply with differences in national and regional legislation Specific details can be found in the delivery documentation or from your dealer or representative Please note that your warranty may be invalidated if this instrument is modified in any way or not used as intended by Owlstone Medical accessories and reagents which have not been approved by Owlstone Medical are used the instrument ...

Page 9: ...isposed of in normal waste Please consult Owlstone Medical Ltd for details of our recycling and disposal program for this product For users outside the European Union consult local authorities for correct disposal or contact Owlstone Medical Ltd The Air Filter contains activated carbon and cannot be recycled as part of your recycling collection scheme nor at household waste recycling centres Used ...

Page 10: ...nalysis It is supplied only on a professional business to business basis for the purpose of research and development Technical Specifications Operating Condition Description Input Voltage 230V AC 50Hz standard 115V AC 60Hz US variant Equipment Rating 350W 230V 1 5A standard 350W 115V 3A US variant Storage Temperature range 20o C 50o C unit off Operating Temperature range 10o C 30o C Operating Humi...

Page 11: ... above the recommended operating temperature range Identification of Parts The CASPER Portable Air Supply is supplied with the following parts 01 0931 CASPER Unit 230V Variant 01 0937 CASPER Unit 115V variant 01 0862 Air Filter 01 0924 Air Supply Tubing 50 1869 Mains Power Cord UK 50 1870 Mains Power Cord EU 50 1871 Mains Power Cord US ...

Page 12: ...gloves when setting up the CASPER Portable Air Supply in order to keep parts as clean as possible 1 Remove CASPER from box and foam packaging and place on a clean dry surface away from open solvent sources Note Please retain all packaging during the warranty period The CASPER Portable Air Supply should not be used if it is visibly damaged ...

Page 13: ...the plastic cover from inside the unit 3 Remove the Air Supply Tubing and the Air Filter from the box Note The use by date on the Air Filter is the date by which the Air Filter should have started to be used The Air Filter and Tubing should not be used if packaging is damaged ...

Page 14: ...0006 004 08 Oct 2019 Page 14 of 25 4 Remove the Air Filter from the foam packaging 5 Remove the Air Filter from the plastic case and place in the CASPER unit Note The Air Filter should be stored in the CASPER unit until end of life ...

Page 15: ... sealed plastic bag and attach to the valve connector of the CASPER unit 7 Connect the Mains Power Cord to the IEC connector on the back of the CASPER unit and plug in to mains power Note Avoid positioning the system in a corner or close to walls which may restrict airflow away from the back of the system ...

Page 16: ...rtable Air Supply should not be used near to sources which can contaminate the air supply such as solvents and biologically hazardous materials 8 Following assembly the complete system must be run continuously for 1 hour to allow it to settle and for the air quality to stabilise ...

Page 17: ... Owlstone Medical Ltd 2019 95 0006 004 08 Oct 2019 Page 17 of 25 4 Operating Instructions Figure 1 CASPER User Interface 1 Attach the Air Supply Tubing to the ReCIVA Breath Sampler ...

Page 18: ... status indicator will flash red a buzzer will briefly sound and the fans will pulse The CASPER Portable Air Supply should not be used if either the buzzer does not sound or the red warning LED does not illuminate The remaining lifetime of the scrubber will be shown by the filter status indicators On first use of the scrubber five green LEDs should be visible ...

Page 19: ...dicator will flash green for ten seconds then stay solid green indicating the system is delivering the required flow and is ready to use Note When the CASPER Portable Air Supply is not in use seal the air supply pipe using the attached bung The CASPER Portable Air Supply should not be used if the LEDs turn off during operation ...

Page 20: ...up Solid green System operating within specified limits for air flow and air temperature Solid amber WARNING System unavailable for use Flashing red ALARM Remove air supply from test subject an audible alarm will be present Once the air supply has been removed from the test subject press the on off switch to turn off the audible alarm ...

Page 21: ... filter life remaining 20 Air filter life remaining 20 Order replacement Air Filter Air filter life remaining 2 hours Air filter expired Air filter not fitted Note At the end of life remove the Air Filter from the system and replace with a new Air Filter Wear clean nitrile gloves when changing the Air Filter The Air Supply Tubing should also be replaced at the end of filter life ...

Page 22: ... switch and try to start pump again If this does not clear the error request service System Indicator Flashing red Audible Alarm Blockage of the Air Supply Tubing while the system is providing air Remove air supply from test subject if present Turn off flow using on off switch If possible remove blockage and reset system by hard power off and on using IEC power inlet rocker switch If not replace A...

Page 23: ...The system has been run for more than 24 hours continuously Remove air supply from test subject if present Turn off flow using on off switch Reset system by hard power off and on using IEC power inlet rocker switch The Air Supply Tubing has been disconnected from the unit while the system is providing air Remove air supply from test subject if present Turn off flow using on off switch Re connect A...

Page 24: ...Indication Possible Cause Action System Indicator Flashing red Audible Alarm The system has detected an internal error while the system is providing air Remove air supply from test subject if present Turn off flow using on off switch Turn off power to system Request service ...

Page 25: ...t or this procedure please contact Owlstone Medical Ltd immediately using the contact details below Support forum https support owlstonenanotech com Email support owlstone com Address Owlstone Medical Ltd Unit 183 Cambridge Science Park Milton Road Cambridge CB4 0GJ United Kingdom Telephone 44 0 1223 428200 8 Cleaning Unplug the system prior to cleaning the device The system should only be cleaned...

Reviews: