MAINTENANCE
Note
Maintenance and inspection procedure must be performed without patient positioned
in the equipment.
This unit, like all other electrical appliances, must be used correctly and also serviced and
controlled at regular intervals. This precaution ensures safe and efficient performance.
Regular maintenance consists of checks performed by the operator and/or by a qualified
technician.
The operator can control the following items:
Frequency
Type of check
Method
Daily
Functioning of the indicator lights
Visual inspection
Daily
Check that the cables do not show signs of
breaking or wear
Visual inspection
Daily
Check that the unit is not damaged externally in
such a way that the safety of protection from
radiation is compromised
Visual inspection
Daily
Check that there are no traces of oil on the tube-
head
Visual inspection
Daily
Check that arm movement is smooth
Practical inspection
Monthly
Integrity of equipment and labels
Visual inspection
Every 6 Months
QC test
refer to paragraph 8.5
Warning
If the operator detects irregularities or failures, he must immediately call Technical
Service.
Besides the above controls, the Service Engineer will also check the following during preventive
maintenance:
Frequency
Type of check
Annually
Correct equipment centring
Annually
Check technical factors
Annually
Perform sensor calibration
Annually
Check that the fixing screws are tightened