18 | Ottobock
2) Follow the instructions for repair provided with the repair kit (see Fig. 7).
3) Re-inflate the ROHO insert with air.
4) Immerse the ROHO insert in water to check that the puncture has been sealed.
Inserting the ROHO insert
1) Dry the ROHO insert thoroughly.
2) Position the ROHO insert so that the valve fits into the valve hole in the foam base (see
Fig. 22).
3) Carefully reinsert the ROHO insert. The edge must be located between the foam and the
black cushion base. Do not pull on the air cells (see Fig. 21).
4) Put the cover back on (see Page 16).
5) Set up the cushion for the user again (see Page 16).
8 Legal Information
8.1 Liability
The manufacturer’s warranty applies only if the device has been used under the conditions and
for the purpose described. The manufacturer recommends that the device be used and main
tained according to the instructions for use.
The manufacturer is not responsible for damages caused by components and spare parts not
approved by the manufacturer. Repairs must be carried out exclusively by authorised dealers or
by the manufacturer.
8.2 CE Conformity
This device meets the requirements of the 93
/
42
/
EEC guidelines for medical devices. This
device has been classified as a class I device according to the classification criteria outlined in
appendix IX of the guidelines. The declaration of conformity was therefore created by Ottobock
with sole responsibility according to appendix VII of the guidelines.
8.3 Trademarks
All denotations within this accompanying document are subject to the provisions of the respective
applicable trademark laws and the rights of the respective owners, with no restrictions.
All brands, trade names or company names may be registered trademarks and are subject to the
rights of the respective owners.
Should trademarks in this accompanying document fail to be explicitly identified as such, this
does not justify the conclusion that the denotation in question is free of third-party rights.
9 Appendices
9.1 Technical data
General information
Cushion height*
100 mm / 4 inches
Weight (420 x 420 mm; 16 x 16 inches)
1200 g
Maximum user weight
150 kg / 330 lbs
Foam base
Material
PUR extra-cushioning foam
Foam density (ISO 845)
100 kg/m
3
Compressive stress of the foam (ISO 3386)
11 kPa
Cover
Material
Dartex
Summary of Contents for Terra Flair
Page 2: ...1 2 3 4 5 6 2 Ottobock...
Page 3: ...7 8 9 10 11 12 Ottobock 3...
Page 4: ...13 14 15 16 17 18 4 Ottobock...
Page 5: ...19 20 21 22 Ottobock 5...
Page 40: ...40 Ottobock Anchura y profundidad del asiento de la silla de ruedas...
Page 41: ...Ottobock 41...
Page 42: ...42 Ottobock...