Kenevo

 

Although this knee is allowed for use with the TaiLor Made foot, Ottobock would not typically suggest this 
K2 knee with a carbon foot except under special circumstances. 

 

Please contact an  Ottobock Professional Clinical  Service clinician  if wanting to use this  combination.

 

 

C-Leg  Bench Alignment Recommendation from  Instructions For Use

 

 

•   Position the  middle  of the  foot (MF) approx. 30  mm/1.18 inch  anterior to the  alignment 

reference line (A). This applies to all foot components that  are  recommended for use with 
the  product, regardless of the  previous alignment specifications in the  Instructions for Use 
of those feet.

 

 

•   Noting  the  alignment recommendation of the  foot component, add 5 mm to the  effective 

heel height (shoe heel  height – sole  thickness in the  forefoot  area), and  set  the  outward 
rotation  of the  foot.

 

 

•   Place the  alignment reference point  (=knee axis) approx. 0-5  mm/0-0.19 inch  anterior to 

the alignment reference line.  Take  into account the  knee-ground distance and 
outward rotation of the knee (the adapter insert provides for a rotation of approx. 5°). 
The recommended sagittal positioning of the  alignment reference point is  20 
mm/0.79 inch  above the  medial tibial plateau.

 

 

•    Connect the foot and knee joint using a tube adapter. To do so, tilt the joint in the correct position 
 

and  set  the  required tube length.

 

 

•   Mark the  lateral  center of the  socket with a centered, proximal  dot  and  a centered, distal 

dot. Mark a line through both  dots from the  socket brim  to the  end of the  socket. Use  the 
4H105 knee extender (see IFU page 22).

 

 

•   Now  position the  socket such that  the  alignment reference  line passes  through the 

proximal centre dot.  Adjust  the  socket flexion to 3° – 5°,  but  take  the  individual  situation 
(e.g. hip  joint contractures) and  the  ischial  tuberosity-to ground distance  into account.

 

 

•   The  adjustment software helps to determine the  socket flexion precisely.

 

 

•   Connect the  socket and  modular knee joint using an  adapter.

 

 

C-Leg  Static Alignment Recommendation from  the  Instructions For Use

 

 

•   To determine the  load  line,  have  the  patient (with shoes) stand on  the  force  measurement 

plate with the  prosthesis side and  on  the  height compensation plate  with the  other leg.  The 
prosthesis side must  be  sufficiently  loaded (> 35% body  weight). Note  the  weight  display on 
the  L.A.S.A.R. Posture.

 

 

•   Optimize the  alignment solely  by changing the  plantar flexion. Only make  adjustments to 

the distal  and  proximal  set  screws of the  socket adapter on  the  prosthetic foot,  so  that  the 
load  line (laser line) runs approx. 30  mm/1.18 inch  in front of the  alignment reference point 
(= knee axis) of the  knee joint.

 

 

 

 

 