10) Apply Loctite to the thread of the mounting screw.
11) Slide the new functional ring onto the adapter of the prosthetic foot. The
structure of the functional ring only permits installation with the correct
alignment.
12) Insert the adapter of the prosthetic foot into the base body and attach it
with the mounting screw (tightening torque:
15 Nm
). To do so, insert the
torque wrench through the hole in the pyramid.
13) Pre-compress the new functional ring (see page 21).
14) Reinstall the prosthetic foot on the prosthesis and pull on the footshell.
8 Disposal
In some jurisdictions it is not permissible to dispose of the product with
unsorted household waste. Improper disposal can be harmful to health and
the environment. Observe the information provided by the responsible
authorities in your country regarding return, collection and disposal proced
ures.
9 Legal information
All legal conditions are subject to the respective national laws of the country
of use and may vary accordingly.
9.1 Liability
The manufacturer will only assume liability if the product is used in accord
ance with the descriptions and instructions provided in this document. The
manufacturer will not assume liability for damage caused by disregarding the
information in this document, particularly due to improper use or unauthor
ised modification of the product.
9.2 CE conformity
The product meets the requirements of Regulation (EU) 2017/745 on medic
al devices. The CE declaration of conformity can be downloaded from the
manufacturer's website.
10 Technical data
F23 Maverick VS
Sizes [cm]
23
24
25
26
27
28
29
30
Heel height [mm]
10
System height [mm]
181
189
202
Build height [mm]
209
217
230
Average weight with footshell [g]
920
1072
1166
Max. body weight [kg]
147
25