Allowable foot size
See item number
in Figure
2
"Technical data" on page 3
3.3 Combination possibilities
CAUTION
Failure to observe the manufacturer requirements regarding
combination possibilities
Injuries, malfunctions or product damage due to unallowable com
bination of prosthesis components
►
Based on the instructions for use of all prosthetic components
used, verify that they may be combined with each other and are
approved for the patient's field of application.
►
Contact the manufacturer with questions.
INFORMATION
In a prosthesis, all prosthetic components have to meet the patient's
requirements regarding the amputation level, body weight, activity
level, environmental conditions and field of application.
3.4 Environmental conditions
CAUTION
Use under unallowable environmental conditions
Fall due to damaged product
►
Do not expose the product to unallowable environmental condi
tions (see the table "Unallowable environmental conditions" in
this section).
►
If the product was exposed to unallowable environmental condi
tions, take suitable steps (e.g. cleaning, repair, replacement,
inspection by the manufacturer or a specialist workshop etc.).
Allowable environmental conditions
Temperature range: -10 °C – +60 °C
20
Allowable environmental conditions
Use - Relative humidity: 0 % – 90 % (Non-condensing)
Storage/Transportation - Relative humidity: 100 %
Tyre pressure: 250 – 1100 mbar
Commercially available, solvent-free cleaning agents
Unallowable environmental conditions
Storage/Transportation: Mechanical vibrations, Impacts
Hygroscopic particles (e.g.Talcum powder), Dust, Sand, Fresh
water, Salt water, Acids, Sweat, Urine
Cleaning agents containing solvents
3.5 Reuse and service life
CAUTION
Reuse on another patient
Fall due to loss of functionality as well as damage to the product
►
Only use the product for a single patient.
CAUTION
Exceeding the service life
Fall due to change in or loss of functionality and damage to the
product
►
Ensure that the service life defined in this section is not
exceeded.
This prosthetic component has been load tested by the manufacturer.
Maximum service life is 2 years.
Summary of Contents for 7E8
Page 11: ...7Z68 5 mm 3 mm 7 6 Prothese fertigstellen Finishing the prosthesis 10 11...
Page 12: ...Lubrication Schmieren 633F16 9 Wartung Maintenance 11 12...
Page 82: ...3 3 1 3 2 2 3 2 3 2 3 3 3 82 3 4 10 C 60 C 0 90 100 250 1100 mbar...
Page 83: ...3 5 2 4 83 5 1 2 6 2 3 SH...
Page 84: ...7 7 1 7 2 3 4 84 4 5 1 Plastilin 636K6 2 699B1 4 cm 3 616G12 6 7...
Page 85: ...0 mm 0 mm 5 6 6 7 7 8 7 3 8 9 85 8 9 7 4 9 10...
Page 86: ...7 5 519L5 7 6 10 11 7Z68 86 8 8 1 8 2 1 2 3 634A58...
Page 87: ...9 30 3 87 11 12 10 11 11 1...
Page 88: ...11 2 CE 93 42 VII 88...
Page 89: ...89...
Page 90: ...90...
Page 91: ...91...