7.2 Standard commands
Command
Response
Description
a!
a<CR><LF>
Acknowledgement active
a
– sensor address; factory setting:
0
aI!
allccccccccmmmmmmvvv
…
…
xxxxxxxxxxxxx<CR><LF>
Send identification
a
– sensor address
ll
– SDI-12 protocol version
cccccccc
– manufacturer’s identification
(company name)
mmmmmm
– sensor identification
vvv
– sensor version (in this case firmware)
xxxxxxxxxxxxx
– additional identification (in this case
serial number; max. 13 characters)
OTT PLS 500 response
014OTTHYDROPLS500100
…
…
xxxxxxxxxx
aAb!
b<CR><LF>
Change sensor address
a
– old sensor address
b
– new sensor address
?!
a<CR><LF>
Query sensor address
a
– sensor address
aV!
atttn<CR><LF>
Perform system test
a
– sensor address
ttt
– time in seconds until the sensor provides the
result of the system test
response OTT PLS 500:
000
n
– number of measured values
response OTT PLS 500:
9
aD0!
a<value1><value2><value3>
…
…
<CR><LF>
Send data (after
aV!
)
a
– sensor address
<value1>
– relative humidity in the probe housing
measured value format:
pbb
[
% rH
]
<value2>
– dew point in the probe housing
measured value format:
pbb.ee
[°C]
<value3>
– current sensor orientation
measured value format:
pbbb
[°]
aD1!
a<value4><value5><value6>
…
…
<CR><LF>
Send data (after
aV!
)
a
– sensor address
<value4>
– sensor orientation during installation
measured value format:
pbbb
[°]
<value5>
– corrected temperature value of pressure sensor
measured value format:
pbb.ee
[°C]
<value6>
– temperature of internal humidity sensor
measured value format:
pbb.ee
[°C]
aD2!
a<value7><value8><value9>
…
…
<CR><LF>
Send data (after
aV!
)
a
– sensor address
<value7>
– corrected pressure value of pressure sensor
measured value format:
pbbbb.ee
[mbar]
<value8>
– standard deviation of corrected pressure value
of pressure sensor within the averaging time
1)
measured value format:
pbbbb.ee
[mbar]
<value9>
– device status
+0
➝
no error occured
+1
➝
system reset – status flag is set when
connecting the operating voltage/upon
unexpected reset; deletion after status
read out
(Continuation of device status see next page)
1)
determined from 2 … 238 single measurements of a measurement interval, (see command
aXXC!
)