5 | Assembly
30
SMP12 Smart Pyranometer | EN
5.2 Installation for measuring global radiation on sloping surfaces
In a photovoltaic system, the pyranometer must be installed at the same zenith angle and azimuth angle as the
modules. The pyranometer can be mounted using the adjustable leveling feet or using a set of fixed feet that
are suitable for mounting on sloping surfaces.
2
1
1
Zenith angle ± 1°
2
Azimuth angle ± 2°
4
Place the pyranometer on a horizontal surface.
4
Make sure that the leveling feet protrude as far as the adjustable foot.
4
Level the pyranometer (see Chapter
4
Label the pyranometer with a note stating that the feet have been set.
4
Alternatively, remove the leveling feet and mount the fixed feet.
4
Label the pyranometer with a note stating that the fixed feet are suitable for sloping installation.
4
Fix the pyranometer on the sloping surface.
Summary of Contents for KIPP & ZONEN SMP12
Page 3: ...US English 4 Deutsch 38 Fran ais 72 Italiano 106 Espa ol 140 174 SMP12 Smart Pyranometer 3...
Page 177: ...1 SMP12 Smart Pyranometer ZH 177 IC ICES 003 B B ICES 003 B NMB 003 B 1 4 i 4 1 4 2 1 2...
Page 178: ...1 178 SMP12 Smart Pyranometer ZH 1 5...
Page 179: ...2 SMP12 Smart Pyranometer ZH 179 2 4 4 2 1 4 4 4 2 2 SMP12 4 SMP12 2 3 4 4 OTT HydroMet...
Page 180: ...2 180 SMP12 Smart Pyranometer ZH 2 4 4 4 4 4 2 5 60 4 2 6 4 4 4 OTT HydroMet 4...
Page 181: ...3 SMP12 Smart Pyranometer ZH 181 3 3 1 SMP12 Modbus RS 485 PLC...
Page 189: ...4 SMP12 Smart Pyranometer ZH 189 4 5 2 Smart Explorer 4 Smart Explorer...
Page 190: ...4 190 SMP12 Smart Pyranometer ZH 4 Setup 4 4 Update...
Page 192: ...4 192 SMP12 Smart Pyranometer ZH 4 4 Confirm 4 5 4 4 Configuration...
Page 196: ...4 196 SMP12 Smart Pyranometer ZH 4 Start From Boot 4 1 Connections 4 Configuration 4 10...
Page 197: ...5 SMP12 Smart Pyranometer ZH 197 5 5 1 5 1 1 5 5 10 h 10 x h 2 360 5 0 8 4 4 100 4...
Page 199: ...5 SMP12 Smart Pyranometer ZH 199 4 50 4 4 4 4 4 4 4...
Page 200: ...5 200 SMP12 Smart Pyranometer ZH 5 2 2 1 1 1 2 2 4 4 4 198 4 4 4 4...
Page 202: ...5 202 SMP12 Smart Pyranometer ZH 5 4 0 1 1 2 3 6 4 5 1 2 3 4 5 6 4 198 4 201...
Page 203: ...5 SMP12 Smart Pyranometer ZH 203 5 5 5 3 1 2 4 3 1 2 3 4 5...
Page 204: ...5 204 SMP12 Smart Pyranometer ZH 4...
Page 206: ...6 206 SMP12 Smart Pyranometer ZH 6 3 ElektroG 2012 19 EU OTT HydroMet 4 4 4...
Page 207: ...7 SMP12 Smart Pyranometer ZH 207 7 7 1 4 4 12 24 V 4 4 4 4 OTT HydroMet...