Quick Guide: 

Bio-logic

®

 NavPRO ONE

®

 ABR screening

Otometrics - Next generation Bio-logic products

4

Bio-logic

®

Select the appropriate ear tip size and place it on the probe tip. Insert the probe in the ear making  
sure that the ear tip fits tightly into the ear canal.

Select QuickABR button

Select the test ear by pressing 

either the Right or Left button

A calibration will be performed 

prior to measurement.  
A green light indicates  

successful calibration. 

An impedance check  

will be performed prior to  

measurement. An impedance 

below 6 k ohm is recommended 

for best test results. 

After successful ear probe  
calibration and impedance 
check, the test will start  
automatically.  If yellow or  
red lights are displayed, check 
to make sure you are using the 
correct probe tip and ear tip 
size, that the eartip is inserted 
correctly, and that there is no 
excessive ambient noise.  

Summary of Contents for Bio-Logic NavPRO ONE

Page 1: ...S 45 45 75 55 55 otoinfo natus com Natus Medical Incorporated 1 800 289 2150 otoorder us natus com www otometrics com Bio logic NavPRO ONE ABR screening Quick start guide Doc No 7 26 1388 EN 00 Part No 7 26 13808 EN Bio logic ...

Page 2: ...Turning On Off and charging the device Start menu On off button on right side Current time Note this screen test options can vary depending on your set up Battery indicator Battery status Select patient Additional information about options for current screen Turn off device Customize Power supply connector ...

Page 3: ... screening For binaural testing instructions please refer to the ABR diagnostic Quick Start Guide Auditory brainstem response abr screening Connect the electrode cable to the white and the probe to the red connector Clean the patient s skin and place the electrode pads Connect the clips to the corresponding sites White high forehead Black cheek Red mastoid ...

Page 4: ...Right or Left button A calibration will be performed prior to measurement A green light indicates successful calibration An impedance check will be performed prior to measurement An impedance below 6 k ohm is recommended for best test results After successful ear probe calibration and impedance check the test will start automatically If yellow or red lights are displayed check to make sure you are...

Page 5: ...wipe screen up and down to change display scale Impedance monitor impedance should be below 6 kOhm for best results Overall screening test result Pause resume test Test parameters and conditions Print results to a label or to a pdf file Normative data for selected age red wave III green wave V Response detection bar Ongoing EEG value should remain in the lower third Test progress Test other ear St...

Page 6: ...serves the right to make changes at any time without notice Please note that not all products displayed on our webpage or advertising materials are available in all territories Please contact your local distributor or PATH medical for further information The latest revision of the Quick Start Guide is available online at www pathme de download Errors and omissions excepted Copyright Notice No part...

Reviews: