otometrics Bio-logic NavPRO ONE Quick Start Manual Download Page 6

Quick Guide: 

Bio-logic

®

 NavPRO ONE

®

 ASSR

Otometrics - Next generation Bio-logic products

6

Bio-logic

®

ASSR test screen 

Frequency being tested 

Intensity level being tested 

Impedance monitor 

Simple calibration mode 

Expert calibration mode 

Progress for each 

tested frequency 

Overall EEG, – value should 

remain in the lower third. Press 

to enable raw EEG view

Test progress 

Stop test  Pause ongoing test

Skip this single test

If you are using an ear probe, a calibration will be performed together with the impedance check. You can 
select the simple calibration or the expert calibration mode in the device set up. 

Summary of Contents for Bio-logic NavPRO ONE

Page 1: ...rk ApS 45 45 75 55 55 otoinfo natus com Natus Medical Incorporated 1 800 289 2150 otoorder us natus com www otometrics com Bio logic NavPRO ONE ASSR Quick Start Guide Doc No 7 26 1389 EN 00 Part No 7 26 13809 EN Bio logic ...

Page 2: ...Turning On Off and charging the device Start menu On off button on right side Current time Note this screen test options can vary depending on your set up Battery status Select patient Additional information about options for current screen Turn off device Customize settings Power supply connector ...

Page 3: ...sert earphones blue connector Probe red connector Second probe for binaural testing blue connector Clean the patient s skin and place the electrode pads Connect the clips to the corresponding sites Recommended montage is Testing Connect the electrode cable to the white and the probe to the red connector White high forehead Black cheek Red mastoid For binaural testing place red electrode on the nap...

Page 4: ...e the parameters select one of the presets and then the tools button at the bottom of the screen Select masking noise monaural testing only Select start minimum and maximum level for adaptive mode 40 Hz awake and adult patients 80 Hz babies and higher frequencies Auto mixed rate optimized for performance Half octave stimuli provide better frequency specificity but longer testtime select frequencie...

Page 5: ...ted if stimulating at low intensity levels only Make sure the patient is sitting or lying down comfortably An impedance check will be performed prior to measurement If impedance is below 6 k ohm the start button will become available You can configure the settings for the test to autostart if the impedance is lower than 4 k ohm If impedance is higher than 6 k ohm but below 12 k ohm a skip button w...

Page 6: ...tion mode Expert calibration mode Progress for each tested frequency Overall EEG value should remain in the lower third Press to enable raw EEG view Test progress Stop test Pause ongoing test Skip this single test If you are using an ear probe a calibration will be performed together with the impedance check You can select the simple calibration or the expert calibration mode in the device set up ...

Page 7: ...ts Note These circles indicate that other screens are available to view Swipe your finger right or left to get other views Print results to a label or to a pdf file Press on a check mark or circle to view details Test conditions Please note that the estimated audiogram is only available in the adaptive test mode ...

Page 8: ... to make changes at any time without notice Please note that not all products displayed on our webpage or advertising materials are available in all territories Please contact your local distributor or PATH medical for further information The latest revision of the Quick Start Guide is available online at www pathme de download Errors and omissions excepted Copyright Notice No part of this manual ...