8. Align the microphones along the loudspeaker axis.
9. Position the reference microphone as close as possible to the front
microphone of the hearing instrument without touching.
10. Close the lid. You are ready to start testing.
3.5
ITE hearing instruments
This procedure applies to any type of custom hearing instruments, including ITE (In The Ear), ITC (In The Canal), CIC (Com-
pletely In the Canal).
Positioning the hearing instrument for directional testing is described in
General guidelines for positioning the hearing
.
1. Wrap a thin roll of acoustic putty around the receiver end of the hearing
instrument.
22
Otometrics - AURICAL HIT
3 Testing hearing instruments