Accuracy of sound level
Entire level range (AC):
125 to 5000 Hz: ±3 dB, 5000 to 12500 Hz: ±5 dB
Entire level range (BC):
250 to 5000 Hz: ±4 dB, 5000 to 8000 Hz: ±5 dB
The reference conditions for the specification of frequency response and sound pressure level depend on the type of audi-
ometer. AURICAL Aud can be calibrated as either a “corrected” (Type AE) or “uncorrected” (Type A) speech audiometer:
Type AE calibration:
•
The output sound pressure level and frequency response are specified in terms of free-field equivalent sound pressure
level.
•
The loudspeaker output is specified as measured under free-field conditions, at 1 m distance, and on the axis of the
loudspeaker.
•
Bone vibrator output is not corrected to obtain a free-field equivalent sound force level; uncorrected output is pro-
duced (please see below under “Type A”).
•
Calibration of speech signals is performed using either a 1 kHz pure tone (earphones) or 1 kHz warble tone (loud-
speakers).
Type A calibration:
•
The output sound pressure level and frequency response are specified in terms of coupler level. See table below for
coupler/ear simulator used.
•
The loudspeaker output is specified as measured under free-field conditions, at 1 m distance, and on the axis of the
loudspeaker.
•
Bone vibrator output is not corrected to obtain a free-field equivalent sound force level; uncorrected output meas-
ured by an artificial mastoid (IEC 60318-6) is produced.
•
Calibration of speech signals is performed using either a 1 kHz pure tone (earphones) or 1 kHz warble tone (loud-
speakers).
Transducer type
Coupler/ear simulator
Supra-aural earphone
IEC 60318-3
HDA200/HDA300
IEC 60318-1
Insert phone
IEC 60318-5
Attenuator
1 or 5 dB step resolution over the entire range.
HL Range
The maximum output levels from AURICAL Aud depend on the actual sensitivity of the individual transducers, and they
will be slightly different for each unit. However, the minimum requirements from IEC and ANSI standards are fulfilled for
all units.
They are specified in the following.
Otometrics - AURICAL Aud
121
13 Technical specifications
Summary of Contents for AURICAL Aud
Page 1: ...AURICAL Aud Reference Manual Doc No 7 50 1030 EN 15 Part No 7 50 10300 EN ...
Page 12: ...1 Introduction to AURICAL Aud 12 Otometrics AURICAL Aud ...
Page 70: ...2 Navigating in the OTOsuite Audiometry Module 70 Otometrics AURICAL Aud ...
Page 100: ...5 Unpacking and Installing 100 Otometrics AURICAL Aud ...
Page 104: ...7 Communicating with the device 104 Otometrics AURICAL Aud ...
Page 114: ...11 Troubleshooting 114 Otometrics AURICAL Aud ...