1. Make sure that AURICAL Aud is connected in OTOsuite. (Select your device in the configuration wizard.)
2. Make sure that the firmware is up-to-date. (Open the Audiometry control panel. If there is new firmware, you will be
prompted to update it.)
3. Close OTOsuite.
4. If possible, place someone (for example, yourself, a client or an assistant) in the intended client location.
5. Make sure that the microphone is placed as described in
Microphone placement for use with RoomTune
.
6. Select RoomTune in the Windows
Start
menu.
The RoomTune application will then run in the background and you can access it from the system tray.
The RoomTune icon in the system tray
9.3
Using RoomTune in a new environment
Before using AURICAL Aud speaker or speakers in a new environment, you must run
Full RoomTune
. You should also use
Full RoomTune
if you make significant changes to the environment, such as moving the position of the speaker or adding a
large number of people to the room.
Once you have done the steps described in
, you can do the following:
1. Click on the RoomTune icon to open the context menu.
2. Click
RoomTune Settings
to open the application window.
Otometrics - AURICAL Aud
107
9 Tuning the AURICAL Aud speaker for use with third-party applications
Summary of Contents for AURICAL Aud
Page 1: ...AURICAL Aud Reference Manual Doc No 7 50 1030 EN 15 Part No 7 50 10300 EN ...
Page 12: ...1 Introduction to AURICAL Aud 12 Otometrics AURICAL Aud ...
Page 70: ...2 Navigating in the OTOsuite Audiometry Module 70 Otometrics AURICAL Aud ...
Page 100: ...5 Unpacking and Installing 100 Otometrics AURICAL Aud ...
Page 104: ...7 Communicating with the device 104 Otometrics AURICAL Aud ...
Page 114: ...11 Troubleshooting 114 Otometrics AURICAL Aud ...