Calibration cont...
11. Replace the sensor rainguard with an Otis OI-410 Calibration Cup.
12. Apply a known calibration gas to the OI-410 Calibration Cup that is attached to the sensor housing.
13. Touch the magnet to
MENU
.
14. The display screen should resemble the following illustration, counting down to begin Auto Cal.
15. Once Auto Cal is complete, the OI-6900 will display the caled “Reading: ___”.
16. Unscrew the OI-410 Calibration Cup.
17. Reattach (screw on) the sensor rainguard to the sensor housing.
18. Once Calibration is complete, proceed to the next step.