16
17
7. Indicator light
The orange indicator light is activated during an adjustment session and enables the status of the sound processor to
be easily identified, notably for people who are not familiar with the functioning of the device (teachers, educators, etc.).
Start-up test
At start-up, the light illuminates twice, and then extinguishes.
Stimulation indication
The light regularly flashes depending on the stimulation rate of the implant.
This light configuration can be activated upon request by your programming clinician.
Programme selection
Each quick press on the programme button is followed by one or more consecutive flashes of the light, indicating the
programme selected (once for P1, twice for P2 etc.).
Detecting Battery discharge
The light flashes for 30 seconds to warn of the low battery life. It flashes again 30 minutes later to indicate the full
discharge of the batteries and the cut-off of the sound processor.
This configuration can be deactivated at your request by your clinician.
Detection of any other errors
In case of a significant error that may alter the stimulation, the sound processor restarts (leading to momentary power
loss) and then the light flashes quickly and extinguishes. If the problem persists, contact your local distributor or the
Oticon Medical after-sales service.