Össur SYMBIONIC LEG 3 Quick Reference Card Download Page 2

TRAINING CHECKLIST

1. Train user on level ground.

2. Train user on ramps/stairs.

Make sure the user knows how to:

1. Charge the unit and recognize low battery level warning.

2. Turn power 

on

/

off

 and unlock the keypad.

3. Adjust the heel height when changing shoes.

CHARGING AND POWER

Insert the charger cable into the charging port on the user interface panel on the knee module.

 

Charging

:

  The yellow light on the knee unit blinks when charging and becomes a solid light when it is 

fully charged. The ankle module does not indicate charging status. 

 

Power

:

  Fully charged battery is sufficient for 16–24 hours of continuous use, depending on activity.

 

Warnings

:

  Auditory warning signals indicate impending power loss.

RELAX MODE

Warning

Indicates

2xBEEP

Approx. 5 hrs left

3xBEEP

Approx. 4 hrs left

4xBEEP

Approx. 3 hrs left

5xBEEP

Approx. 2 hrs left, 

Ankle will return to neutral (last set heel height) 

Yellow LED is solid ON

1x long high 

pich beep

Symbionic Leg will turn off

Relax Mode

Hold

Press

Beep

Relax Mode  OFF

 5×

Relax Mode  ON

Chair Exit Mode only